Sample Preparation, Consumables, LC/MS, LC/MS/MS, LC columns, LC/QQQ
IndustriesClinical Research
ManufacturerThermo Fisher Scientific
Importance of the Topic
Rosuvastatin is a potent cholesterol‐lowering agent widely prescribed to reduce cardiovascular risk. Accurate quantification of this drug in human plasma is essential for pharmacokinetic studies, therapeutic drug monitoring, and bioequivalence assessments. Developing a robust analytical workflow ensures high confidence in clinical decision‐making and regulatory compliance.
Objectives and Study Overview
The primary aim of this study was to establish and validate a sensitive, rapid, and reliable SPE‐LC‐MS/MS method for measuring rosuvastatin concentrations in human plasma. By combining Thermo Scientific SOLA solid‐phase extraction with Accucore RP‐MS chromatography and triple quadrupole detection, the authors sought to demonstrate linearity over a wide dynamic range, high recovery, minimal matrix effects, and suitability for routine high‐throughput applications.
Methodology and Instrumentation
Sample Preparation and Extraction
- Plasma samples (100 µL) were processed using a Thermo Scientific SOLA 96-well SPE plate under positive pressure.
- The plate was conditioned with methanol and water, followed by loading of plasma, aqueous formic acid wash, low‐organic wash, and elution with 90 % methanol.
- Eluates were evaporated under nitrogen at 40 °C and reconstituted in 25 % methanol aqueous solution (200 µL).
Chromatography and Mass Spectrometry
- LC separation employed an Accela 600 system and Accucore RP-MS column (2.6 µm, 50 × 2.1 mm) with a matching guard cartridge.
- Gradient elution from 5 % to 95 % methanol (each containing 0.1 % formic acid) over 1 min, total run time 3.6 min; flow rate 0.75 mL/min; column at 60 °C.
- Detection used a Thermo Scientific TSQ Vantage triple quadrupole with heated electrospray ionisation (HESI) in positive mode, monitoring SRM transitions 482.3→258.2 for rosuvastatin and 488.2→264.2 for the d6 internal standard.
Main Results and Discussion
Linearity and Sensitivity
- The method exhibited a linear response (1–1000 ng/mL) with r² = 0.9984 using 1/x weighting.
- Lower limit of quantification was demonstrated at 1 ng/mL with acceptable precision and accuracy.
- Quality control samples at 3, 400, and 750 ng/mL showed intra‐ and inter-day precision (%RSD) below 6 % and accuracy within ±11 %.
- Absolute recovery of rosuvastatin averaged 99.3 % with %RSD of 4.9 % across QC levels.
- Post-extraction spiking experiments revealed negligible matrix suppression or enhancement (<5.3 %).
- Chromatographic separation resolved an endogenous plasma impurity (retention at 1.92 min) from the analyte peak at 1.49 min, ensuring specificity.
Benefits and Practical Applications
This SPE-LC-MS/MS workflow offers:
- High throughput compatibility for clinical and bioanalytical laboratories.
- Minimal solvent and sample consumption, reducing cost and environmental impact.
- Robust performance with excellent reproducibility, facilitating large pharmacokinetic or therapeutic monitoring studies.
Future Trends and Opportunities
Further developments may include:
- Automation of sample preparation to increase throughput and consistency.
- Extension to other statins or related compounds in multi-analyte bioassays.
- Integration with high-resolution mass spectrometry for metabolite profiling.
- Application to alternative biological matrices such as dried blood spots or cerebrospinal fluid.
Conclusion
A rapid, sensitive, and reliable SPE-LC-MS/MS method was fully validated for rosuvastatin quantification in human plasma. The approach combines efficient SOLA extraction and high-performance Accucore RP-MS chromatography with SRM detection, delivering excellent linearity, recovery, precision, and minimal matrix interference. The protocol is well suited for routine clinical studies and pharmaceutical research.
Reference
- Pelat M.; Dessy C.; Massion P.; Desager J.-P.; Feron O.; Balligand J.-L. Journal of Circulation 2003, 107, 2480.
- Trivedi R.K.; Kallem R.R.; Mullangi R.; Srinivas N.R. Journal of Pharmaceutical and Biomedical Analysis 2005, 39(3–4), 661–669.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.