Automated Extraction of a Drugs of Abuse Panel from Human Urine Using Biotage Extrahera LV-200 and Microelution SPE Prior to UPLC-MS/MS Analysis

Applications | 2021 | BiotageInstrumentation
Sample Preparation, LC/MS, LC/MS/MS, LC/QQQ
Industries
Forensics
Manufacturer
Biotage

Significance of the Topic


The reliable detection of a broad spectrum of drugs of abuse in urine is critical in clinical toxicology, forensic investigations and workplace testing. Automated microelution SPE coupled with UHPLC-MS/MS improves sample throughput, reduces solvent consumption and minimizes variability, addressing key challenges in high-volume laboratories.

Objectives and Overview


This study aimed to develop and validate an automated extraction workflow for a multi-class panel of illicit and prescription drugs in hydrolyzed human urine. It compared manual and automated processing using mixed-mode strong cation exchange microelution plates on the Biotage Extrahera LV-200, evaluating recovery, precision, linearity and limits of quantitation.

Methodology


One-milliliter urine samples were spiked with deuterated internal standards and hydrolyzed with β-glucuronidase at 60 °C for 2 h. Samples were diluted in ammonium acetate buffer (pH 5) and loaded onto 2 mg mixed-mode SCX microelution plates. The SPE protocol included conditioning (methanol), equilibration (4 % phosphoric acid), sequential washes (phosphoric acid and 50:50 water/methanol) and elution with dichloromethane/methanol/ammonium hydroxide (78:20:2). Eluates were evaporated, reconstituted in water/methanol (90:10, 0.1 % formic acid) and analyzed by UHPLC-MS/MS.

Instrumentation


  • Automated SPE: Biotage Extrahera LV-200
  • Manual SPE: Biotage PRESSURE+ 96 manifold
  • Evaporation: Biotage SPE Dry concentrator
  • UHPLC: Shimadzu Nexera with Restek Raptor Biphenyl column (100 × 2.1 mm, 2.7 µm)
  • MS/MS: Shimadzu 8060 Triple Quadrupole with ESI

Results and Discussion


Both workflows produced mean recoveries above 60 % with RSDs below 5 % (n=7) across more than 30 analytes. Calibration curves spanning 1–1000 pg/mL achieved R² > 0.999. Lower limits of quantitation ranged from 1 to 250 pg/mL. Automated extraction of 96 samples required approximately 25 minutes (excluding hydrolysis and evaporation), matching or exceeding the performance of manual SPE in reproducibility and throughput.

Benefits and Practical Applications


  • High sample throughput with low solvent and consumable usage
  • Robust, reproducible extraction for diverse drug panels
  • Reduced hands-on time and operator variability
  • Scalable solution for routine clinical, forensic and QA/QC laboratories

Future Trends and Possibilities


Emerging advancements may include further SPE miniaturization, direct on-line coupling to LC-MS systems, expanded target lists for novel psychoactive substances and AI-driven workflow optimization. Green chemistry approaches and integrated robotic platforms will continue to enhance efficiency and sustainability.

Conclusion


The automated mixed-mode SCX microelution SPE method implemented on the Biotage Extrahera LV-200 delivers consistent, high-throughput extraction of a comprehensive drugs of abuse panel in urine, with excellent recoveries, precision and sensitivity, streamlining laboratory operations.

References


  • Biotage AB. Automated Extraction of Drugs of Abuse from Urine Using Biotage Extrahera LV-200 and Microelution SPE Prior to UPLC-MS/MS Analysis. Application Note AN964; 2021.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Automated Extraction of a Drugs of Abuse Panel from Human Urine Using Biotage® Extrahera™ LV-200 and Microelution SPE Prior to UPLC-MS/MS Analysis
Streamlined Sample Preparation of a Drugs of Abuse Panel in Human Nails Using ISOLUTE® SLE+ Prior to UPLC-MS/MS Analysis
Extraction of a Drugs of Abuse Panel from Human Urine Using Biotage® Mikro ABN SPE Microelution Plates Prior to UPLC-MS/MS Analysis