Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE® SLE+ Prior to LC/MS-MS Analysis

Applications | 2018 | BiotageInstrumentation
Sample Preparation, Consumables, LC/MS, LC/MS/MS, LC/QQQ
Industries
Clinical Research
Manufacturer
Shimadzu, Biotage

Significance of the Topic


Steroid hormone profiling in human serum is essential for clinical diagnostics and research applications. The complex matrix and low analyte concentrations require efficient sample preparation to ensure accurate sensitive quantitation by LC MS MS.

Objectives and Study Overview


This study presents a comprehensive workflow using ISOLUTE SLE+ supported liquid extraction plates to isolate a panel of over 20 steroid hormones from human serum. The goals include evaluating recoveries reproducibility linearity and phospholipid removal using two elution solvent systems.

Methodology


Sample pretreatment involves spiking serum with methanolic internal standards. Extracts are processed on ISOLUTE SLE+ 400 µL plates under low pressure equilibrated and eluted with ethyl acetate or ethyl acetate hexane (75:25). Eluates are dried at 40 °C under a stream of air or nitrogen and reconstituted in a 50:50 mixture of mobile phases.
  • Sample loading volume 300 µL (up to 350 µL for increased sensitivity)
  • Elution solvents EtOAc or EtOAc Hex 75/25 (2×500 µL)
  • Drying 20 minutes at 40 °C
  • Reconstitution volume 200 µL

Instrumentation


  • ISOLUTE SLE+ 400 µL supported liquid extraction plate
  • Biotage PRESSURE+96 positive pressure manifold
  • Biotage SPE Dry 96 sample evaporator
  • Shimadzu Nexera UHPLC with ACE C18 column (100×2.1 mm 1.7 µm)
  • Shimadzu 8060 triple quadrupole mass spectrometer with ESI

Main Results and Discussion


The optimized protocols yielded analyte recoveries above 75% (above 80% for most steroids excluding DHEAS) with RSDs below 10%. Phospholipid interferences were effectively removed compared to protein precipitation. Linearity was demonstrated over 5–5000 pg/mL with r2 values above 0.99 and LOQs as low as <5 pg/mL for key analytes.

Benefits and Practical Applications


The SLE+ workflow simplifies sample preparation by eliminating emulsions delivering clean extracts and robust quantitation for bioanalytical laboratories. The method supports high throughput and reliable steroid hormone analysis for clinical research and quality control.

Future Trends and Opportunities


Further automation of the SLE+ procedure and integration with high resolution mass spectrometry may expand steroid panels and enhance sensitivity. Miniaturization of sample volumes and application to other biological matrices are promising directions.

Conclusion


The presented ISOLUTE SLE+ method offers a reproducible and efficient approach for extracting a broad steroid panel from human serum achieving high recoveries excellent linearity and effective matrix cleanup prior to LC MS MS analysis.

References


Biotage (2018) Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE SLE+ Application Note AN890.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Comparison of Sample Preparation Options for the Extraction of a Panel of Endogenous Steroids from Serum Prior to UHPLC-MS/MS Analysis
Quantitative Analysis of Steroid Hormones Using Triple Quadrupole LC/MS/MS
Extraction of a Comprehensive Steroid Panel from Horse Hair Using ISOLUTE® SLE+ Prior to LC/MS-MS Analysis
Extraction of a Comprehensive Steroid Panel from Human Urine Using EVOLUTE® EXPRESS ABN Prior to LC/MS-MS Analysis