HPLC
IndustriesPharma & Biopharma
ManufacturerThermo Fisher Scientific
Significance of the Topic
Black cohosh is widely used to alleviate menopausal symptoms and its active compounds are triterpene glycosides that lack strong UV chromophores
Study Objectives and Overview
This application study compares charged aerosol detection with traditional evaporative light scattering detection for quantifying triterpene glycosides in black cohosh extract
Methodology and Instrumentation
Sample Preparation
- Standard solution of 27-deoxyactein prepared in methanol with serial dilutions from 10 to 100 micrograms per milliliter
- Sample extract dissolved in methanol by sonication, filtered through 0.22 micron PTFE filters
Chromatographic Conditions
- Column Fused Core C18 4.6 x 150 mm 2.7 micron
- Mobile phase A 0.1 percent formic acid in water B acetonitrile gradient from 30 to 60 percent B over 36 minutes
- Flow rate 1.0 milliliter per minute, column temperature 35 degrees Celsius, injection volume 10 microliters
Instrumentation
- HPLC system with quaternary pump and autosampler
- Charged aerosol detector for improved sensitivity and linearity
- Evaporative light scattering detector operated for comparison
Results and Discussion
Calibration Performance
- Charged aerosol detection produced a linear calibration curve for 27-deoxyactein with high correlation
- ELSD required a logarithmic calibration and showed greater variance at low and high concentrations
- Signal to noise improved by approximately tenfold with charged aerosol detection
Sample Analysis
- Eleven triterpene glycosides quantified using relative retention times against the 27-deoxyactein standard
- Total glycoside content in the extract measured at 3.27 percent
Benefits and Practical Applications
Charged aerosol detection offers mass based response nearly independent of analyte structure and enables consistent quantitation of nonchromophoric glycosides in quality control of nutraceuticals
Future Trends and Opportunities
Expansion of charged aerosol detection to other natural product classes and adoption of inverse gradient techniques to correct for solvent composition effects will further improve robustness and throughput in industry and research labs
Conclusion
Charged aerosol detection outperforms ELSD for triterpene glycoside analysis by delivering linear calibration, enhanced sensitivity and reproducibility making it a superior choice for QC applications
References
1. NSF International Black Cohosh Assay by ELSD INA Method 113.001
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.