Determination of Impurities in Pharmaceuticals

Applications | 2021 | ShimadzuInstrumentation
Consumables, HPLC, LC columns
Industries
Pharma & Biopharma
Manufacturer
Shimadzu

Importance of the topic


Pharmaceutical quality control requires reliable detection of trace impurities in active drug substances to ensure safety and efficacy. Famotidine, a widely used histamine‐2 receptor antagonist, must meet stringent pharmacopoeial standards. Implementing a robust analytical method for impurity profiling supports regulatory compliance and safeguards patient health.

Study objectives and overview


This application note describes the development of a reversed‐phase HPLC procedure for quantifying famotidine and related impurities. The aim was to achieve baseline separation, adequate sensitivity, and reproducible retention, in line with Japanese Pharmacopoeia (JP) requirements.

Methodology and instrumentation


The following setup was employed:
  • Column: Shim‐pack™ VP‐ODS, 150 mm × 4.6 mm I.D., 5 µm
  • Mobile phase: 2 g sodium 1‐heptane sulfonate in 900 mL water, pH 3.0 (acetic acid), diluted to 1000 mL; mixed with 240 mL acetonitrile and 40 mL methanol
  • Flow rate: 0.45 mL/min, adjusted to yield a ~6 min retention for famotidine
  • Column temperature: 25 °C
  • Detector: SPD‐M20A UV at 254 nm
  • Injection volume: 5 µL

Main results and discussion


The method provided sharp, symmetric peaks for famotidine and its impurities, with consistent retention time and minimal baseline noise. Selectivity was sufficient to resolve all targets within a reasonable run time. System suitability tests confirmed reproducibility of retention times and peak areas.

Benefits and practical applications


  • Compliance with JP impurity guidelines
  • High reproducibility and routine operation stability
  • Compatibility with standard HPLC equipment in QC laboratories

Future trends and potential uses


Further enhancements may include adoption of ultrahigh‐pressure LC for faster throughput, integration with mass spectrometric detection for structural elucidation of unknown impurities, and exploration of greener solvent systems to reduce environmental impact.

Conclusion


The described HPLC method using Shim‐pack VP‐ODS and a carefully optimized mobile phase delivers a reliable, easy‐to‐implement solution for routine quality control of famotidine impurities.

Reference


Application News L461, Shimadzu Corporation, First Edition: Dec. 2021

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Analysis of Temozolomide
Analysis of Temozolomide
2021|Shimadzu|Applications
Analysis of Oxybutynin
Analysis of Oxybutynin
2021|Shimadzu|Applications
Analysis of Fenofibrate within JP method
Preparative Purification of Ibuprofen and Its Related Substance