Consumables, HPLC, LC columns
IndustriesPharma & Biopharma
ManufacturerShimadzu
Significance of the Topic
Accurate analysis of Diclofenac Sodium, a widely used nonsteroidal anti-inflammatory drug, is essential for ensuring pharmaceutical quality and monitoring environmental residues.Objectives and Study Overview
This work compares conventional HPLC and high-speed UHPLC workflows on Shim-pack GIST C18 columns to optimize assay speed and maintain analytical performance for Diclofenac Sodium quantification.Methodology and Instrumentation
- Chromatographic system: Shimadzu Nexera-i MT
- Columns:
- Shim-pack GIST C18 (250×4.6 mm, 5 µm) for standard HPLC
- Shim-pack GIST C18 (150×4.6 mm, 3 µm) for UHPLC
- Mobile phase: Sodium phosphate buffer pH 2.5 / Methanol (34/66 v/v)
- Flow rates: 0.8 mL/min (HPLC), 1.4 mL/min (UHPLC)
- Column temperature: 25 °C
- Injection volume: 20 µL
- Detection: UV at 254 nm
Main Results and Discussion
The standard HPLC method provided reliable separation but required longer run times. Transitioning to UHPLC with a shorter column and smaller particle size nearly halved analysis time while preserving peak resolution and sensitivity. Comparative chromatograms confirm enhanced throughput without performance loss.Benefits and Practical Applications
- Increased laboratory efficiency through reduced analysis time
- Lower solvent consumption and operational costs
- Applicable to pharmaceutical QC and environmental monitoring
Future Trends and Potential Applications
Further developments in column chemistry and high-pressure systems will drive faster assays. Coupling UHPLC with automated sample handling and data processing will streamline workflows. The approach can be extended to multi-residue pharmaceutical screening in environmental samples.Conclusion
Adopting UHPLC on Shim-pack GIST C18 columns offers a rapid and robust solution for Diclofenac Sodium analysis, boosting throughput and sustainability without compromising data quality.References
- Shimadzu Corporation (2021) Application News L518: Analysis of Diclofenac Sodium using Shim-pack GIST C18.