Extended linear dynamic range with the XLXR detector system on the Thermo Scientific ISQ 7610 single quadrupole GC-MS

Technical notes | 2021 | Thermo Fisher ScientificInstrumentation
GC/MSD, GC/SQ
Industries
Manufacturer
Thermo Fisher Scientific

Importance of the topic


Analytical laboratories in food, environmental, pharmaceutical, petrochemical and clinical sectors frequently encounter samples containing analytes at both trace and high concentrations. Traditional GC-MS detectors with limited linear dynamic range require separate methods or repeated dilutions, reducing throughput and risking inaccurate quantification or misidentification. An extended dynamic range detector can streamline workflows, improve spectral fidelity and accelerate data delivery.

Objectives and study overview


This study evaluates the new Thermo Scientific XLXR detector against the previous-generation electron multiplier on the ISQ 7610 single quadrupole GC-MS. Two investigations were performed:
  • Sensitivity and linearity in selected ion monitoring (SIM) of octafluoronaphthalene (OFN) from 4.3 fg to 43 ng on-column.
  • Full-scan spectral fidelity and isotopic accuracy of OFN at 0.43 ng to 43 ng on-column.

Methodology and experimental design


Both detectors were tested on identical GC-MS configurations without changing any parameters except the detector. The SIM experiment monitored m/z 272 for OFN across seven orders of magnitude. The full-scan experiment evaluated mass spectral quality and isotopic ratios at high concentrations to assess saturation effects.

Použitá instrumentace


  • GC system: Thermo Scientific Trace 1610 with AI/AS 1610 autosampler
  • Column: TraceGOLD TG-SQC (15 m×0.25 mm×0.25 µm) for SIM; TG-5 ms (30 m×0.25 mm×1 µm) for full scan
  • Injection: 1 µL, splitless, 220 °C inlet, 1.2 mL/min He carrier
  • MS system: ISQ 7610 single quadrupole, EI 70 eV, ExtractaBrite source, transfer line 250 °C
  • Detectors: Standard electron multiplier vs XLXR detector, detector gain 3E+05

Key results and discussion


In SIM mode the XLXR detector achieved seven orders of linearity with R² = 0.9995 and residual RSD 5.4%, compared to R² = 0.994 and RSD 17.9% for the standard detector. At high concentration (0.43–43 ng) the previous detector required a quadratic fit due to saturation, while the XLXR maintained linearity (R² = 0.9997, RSD 2.5%). Full-scan spectra at 43 ng showed the XLXR preserved correct ion abundances and library match quality, whereas the standard detector suffered distorted fragmentation and deviated isotopic ratios (up to –5 percentage points).

Benefits and practical applications


The XLXR detector enables:
  • Consolidation of trace and high-level analyses in a single method
  • Reduction of sample dilutions and re-analyses
  • Straightforward quantification via linear regression
  • Improved spectral fidelity for confident compound identification

Future trends and potential uses


Wider adoption of extended-range detectors will support high-throughput screening, comprehensive unknown screening and enhanced automation in QC/QA laboratories. Continued detector innovations could further expand dynamic range, reduce maintenance and integrate advanced data processing for real-time quality control.

Conclusion


The XLXR detector on the ISQ 7610 GC-MS demonstrates a 2–4× improvement in dynamic range and superior spectral fidelity compared to the legacy electron multiplier. These advances allow laboratories to maximize instrument utilization, accelerate data turnaround and maintain confidence in quantitative and qualitative results.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Extended Linear Dynamic range with a new electron multiplier system on Single Quadrupole GC-MS
Thermo Scientific ISQ 7610 Single Quadrupole GC-MS System
Thermo Scientific ISQ 7610 Single Quadrupole GC-MS System
2022|Thermo Fisher Scientific|Brochures and specifications
Determination of the instrument detection limit of the ISQ 7610 single quadrupole GC-MS with the ExtractaBrite electron ionization source
Analysis of aromatics in gasoline by ASTM D5769 using gas chromatography–single quadrupole mass spectrometry