Quantification of Liquid Hydrocarbons with Large Boiling Point Range using the Polyarc Reactor

Applications | 2017 | ARCInstrumentation
GC
Industries
Energy & Chemicals
Manufacturer
Agilent Technologies, ARC

Significance of the topic



Accurate quantification of complex hydrocarbon mixtures across a wide boiling point range is essential in many fields including petrochemical analysis, environmental monitoring, and quality control. Traditional gas chromatography–flame ionization detection (GC-FID) requires calibration curves for each compound, which is time-consuming and may introduce errors when mixtures differ from standards. The Polyarc® reactor offers a solution by converting all hydrocarbons to methane, yielding a uniform FID response per carbon atom and eliminating the need for individual calibration standards.

Objectives and study overview



This study evaluates the performance of a GC system equipped with a Polyarc® reactor for quantifying C5 to C18 n-alkanes without calibration standards. Key aims include assessing accuracy, separation efficiency, and the reactor’s ability to detect and correct common GC inlet issues such as preferential vaporization and leaks.

Methodology



A mixture of 13 n-alkanes (C5–C18) spanning 2.4 to 21.1 wt% was injected in splitless mode onto an Agilent 7890A GC system. The sample passed through either a traditional FID or a Polyarc®/FID configuration. N-decane served as an internal standard. The injector and column conditions were optimized to maintain constant inlet pressure and sufficient flow to prevent discrimination of volatile components.

Instrumental setup



  • Gas chromatograph: Agilent 7890A with split/splitless inlet
  • Reactor: Polyarc® ARC PA-RRC-A02 with manual flow module PA-CAS-A07
  • Carrier gas: Helium (99.999%)
  • FID gases: Hydrogen (99.999%) and zero-grade air
  • Column: Rtx-2887, 10 m × 0.53 mm, 2.65 µm film thickness
  • Inlet temperature: 360 °C; oven ramp: 35 °C to 245 °C at 15 °C/min

Main results and discussion



The Polyarc®/FID system achieved an average quantification error of 0.9% and a maximum error of 2.4% relative to known concentrations. Peak width (FWHM) increased by only 9% compared to FID-only analysis, demonstrating minimal loss of separation efficiency. Experiments at reduced inlet pressure (3.33 psi vs. 6.64 psi) revealed systematic under-reporting of low-boiling analytes due to preferential vaporization. Standard mixture analysis with Polyarc® allowed identification and correction of injection conditions and detection of column fitting leaks.

Benefits and practical applications



  • Eliminates the need for compound-specific calibration standards
  • Delivers uniform detector response per carbon atom for diverse hydrocarbons
  • Simplifies routine quantification and reduces analysis time
  • Enables detection of inlet and column issues such as discrimination and leaks
  • Applicable in petrochemical quality control, environmental analysis, and process monitoring

Future trends and opportunities



Future developments may extend Polyarc® technology to more complex mixtures including aromatic and heteroatom-containing compounds. Integration with online and real-time monitoring platforms can further enhance process analytics. Continued refinement of reactor design and flow control will broaden the applicability to high-throughput industrial laboratories.

Conclusion



The Polyarc® reactor coupled with GC-FID provides a robust, accurate, and efficient method for quantifying hydrocarbons over a wide boiling range without calibration standards. Its ability to detect and correct inlet artifacts further improves data reliability, making it a valuable tool across analytical laboratories.

References



  1. Scanlon JT, Willis DE. Journal of Chromatographic Science, 23 (1985) 333–340.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor
GC/ FID Analysis of Fatty Acid Methyl Esters without Correction Factors Using the Polyarc Reactor
Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC Polyarc/FID
Enhanced Chromatographic Analysis with Polyarc Ultra
Enhanced Chromatographic Analysis with Polyarc Ultra
2019|Agilent Technologies|Technical notes