Solid Core 4 μm HPLC Column Comparison to Fully Porous 3 μm and 5 μm Columns: Efficiency and Pressure

Technical notes | 2012 | Thermo Fisher ScientificInstrumentation
Consumables, LC columns
Industries
Manufacturer
Thermo Fisher Scientific

Significance of the Topic


Solid core (superficially porous) particles represent an important innovation in HPLC, combining the efficiency advantages of sub-2 µm materials with moderate backpressure, enabling faster, high-resolution analyses on conventional systems.

Objectives and Study Overview


  • Compare chromatographic performance of 4 µm solid core columns to 5 µm and 3 µm fully porous columns.
  • Evaluate pressure drop, separation efficiency, and impedance across a range of flow rates.

Methodology


  • Columns: Thermo Scientific Accucore XL 4 µm (solid core), fully porous 5 µm and 3 µm, all 150 × 4.6 mm.
  • Mobile phase: water/acetonitrile (50 : 50 v/v); column temperature: 30 °C.
  • Flow rates from 200 to 2000 µL/min; test compounds: o-xylene and theophylline (t₀ marker).
  • Data analysis: van Deemter plots for efficiency; kinetic plots for efficiency per unit time and impedance.

Instrumentation


  • High-performance liquid chromatography systems with UV detection or MS interfaces (ESI/APCI).
  • Thermo Scientific Accucore XL column hardware as the separation medium.

Main Results and Discussion


  • Pressure: Accucore XL 4 µm exhibits ~42 % higher backpressure than 5 µm but ~13 % lower than 3 µm across tested flows; at 2 mL/min, pressure reaches ~200 bar.
  • Efficiency: 75 % increase over 5 µm and 50 % over 3 µm fully porous media, with a flatter van Deemter profile allowing a wide operational range.
  • Impedance: Efficiency per unit time improved by 59 % versus 5 µm and 53 % versus 3 µm; overall impedance reduced by 79 % and 72 % respectively, supporting faster, sharper separations at lower backpressure.

Benefits and Practical Applications


  • Enhanced resolution and sensitivity on conventional HPLC platforms without system modifications.
  • Balanced performance enabling high-throughput QA/QC, pharmaceutical, and industrial analyses.
  • Flexible method development due to broad optimal flow rate range.

Future Trends and Applications


  • Broader integration of superficially porous particles in UHPLC and longer column formats.
  • Advancements in column coatings for targeted separations (e.g., chiral, biomolecules).
  • Optimization of system dead volume and source components to fully exploit low impedance.
  • Expansion into automated, high-throughput workflows coupling HPLC with advanced MS detection.

Conclusion


Thermo Scientific Accucore XL 4 µm solid core HPLC columns provide a superior balance of high efficiency and manageable backpressure compared to fully porous 5 µm and 3 µm columns, delivering faster separations, sharper peaks, and improved overall performance in routine analytical workflows.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter
Thermo Solid Core Technical Compendium
Thermo Solid Core Technical Compendium
2012|Thermo Fisher Scientific|Guides
Comparison of Solid Core HPLC Column Performance
Comparison of Solid Core HPLC Column Performance
2012|Thermo Fisher Scientific|Technical notes
Thermo Scientific Accucore XL HPLC Columns
Thermo Scientific Accucore XL HPLC Columns
2012|Thermo Fisher Scientific|Brochures and specifications