Consumables, LC columns
IndustriesManufacturerThermo Fisher Scientific
Significance of the Topic
Solid core (superficially porous) particles represent an important innovation in HPLC, combining the efficiency advantages of sub-2 µm materials with moderate backpressure, enabling faster, high-resolution analyses on conventional systems.
Objectives and Study Overview
- Compare chromatographic performance of 4 µm solid core columns to 5 µm and 3 µm fully porous columns.
- Evaluate pressure drop, separation efficiency, and impedance across a range of flow rates.
Methodology
- Columns: Thermo Scientific Accucore XL 4 µm (solid core), fully porous 5 µm and 3 µm, all 150 × 4.6 mm.
- Mobile phase: water/acetonitrile (50 : 50 v/v); column temperature: 30 °C.
- Flow rates from 200 to 2000 µL/min; test compounds: o-xylene and theophylline (t₀ marker).
- Data analysis: van Deemter plots for efficiency; kinetic plots for efficiency per unit time and impedance.
Instrumentation
- High-performance liquid chromatography systems with UV detection or MS interfaces (ESI/APCI).
- Thermo Scientific Accucore XL column hardware as the separation medium.
Main Results and Discussion
- Pressure: Accucore XL 4 µm exhibits ~42 % higher backpressure than 5 µm but ~13 % lower than 3 µm across tested flows; at 2 mL/min, pressure reaches ~200 bar.
- Efficiency: 75 % increase over 5 µm and 50 % over 3 µm fully porous media, with a flatter van Deemter profile allowing a wide operational range.
- Impedance: Efficiency per unit time improved by 59 % versus 5 µm and 53 % versus 3 µm; overall impedance reduced by 79 % and 72 % respectively, supporting faster, sharper separations at lower backpressure.
Benefits and Practical Applications
- Enhanced resolution and sensitivity on conventional HPLC platforms without system modifications.
- Balanced performance enabling high-throughput QA/QC, pharmaceutical, and industrial analyses.
- Flexible method development due to broad optimal flow rate range.
Future Trends and Applications
- Broader integration of superficially porous particles in UHPLC and longer column formats.
- Advancements in column coatings for targeted separations (e.g., chiral, biomolecules).
- Optimization of system dead volume and source components to fully exploit low impedance.
- Expansion into automated, high-throughput workflows coupling HPLC with advanced MS detection.
Conclusion
Thermo Scientific Accucore XL 4 µm solid core HPLC columns provide a superior balance of high efficiency and manageable backpressure compared to fully porous 5 µm and 3 µm columns, delivering faster separations, sharper peaks, and improved overall performance in routine analytical workflows.