Fast Analysis of Ketones Using a Core Enhanced Technology Accucore HPLC Column

Applications | 2011 | Thermo Fisher ScientificInstrumentation
Consumables, HPLC, LC columns
Industries
Energy & Chemicals
Manufacturer
Thermo Fisher Scientific

Significance of the Topic


Efficient separation and quantification of ketones play a pivotal role in industries such as pharmaceuticals, petrochemicals and food safety. High-throughput HPLC methods enable rapid quality control, reduce analysis times and support fast decision-making in research and production environments.

Goals and Study Overview


This application note evaluates the performance of a Core Enhanced Technology Accucore C18 column (2.6 µm, 50 × 2.1 mm) for ultra-fast ketone analysis and compares it with a conventional Hypersil GOLD column (5 µm, 150 × 4.6 mm). The primary aim is to demonstrate a significant reduction in cycle time while maintaining resolution and peak shape for a mixture of seven ketones.

Methodology and Instrumentation


Sample Preparation:
  • Standard ketone mixture prepared in a 1:1 ratio of acetonitrile and water.
Chromatographic Conditions:
  • Instrument: Thermo Scientific Accela HPLC/UHPLC.
  • Column: Accucore C18, 2.6 µm, 50 × 2.1 mm; Hypersil GOLD, 5 µm, 150 × 4.6 mm.
  • Column temperature: 45 °C.
  • Injection volume: 1 µL.
  • Flow rate: 1.0 mL/min.
  • Detection: PDA/UV at 258 nm.
  • Mobile phase A: Water; Mobile phase B: Acetonitrile.
  • Gradient program adjusted for each column to optimize separation within targeted cycle times.
Consumables:
  • HPLC-grade water and acetonitrile.
  • Certified vials with PTFE-lined caps.

Main Results and Discussion


The Accucore C18 column achieved baseline separation of seven ketones in under 1 minute, compared to an 11 minute run on the Hypersil GOLD. This represents a seven-fold increase in throughput. Key observations include:
  • Retention times on Accucore ranged from 0.18 to 0.73 min versus 2.68 to 4.95 min on Hypersil GOLD.
  • Sharp, symmetrical peaks with minimal tailing due to reduced secondary interactions on the core–shell stationary phase.
  • Back-pressure remained below 330 bar, compatible with standard HPLC systems.

Benefits and Practical Applications


The fast method provides:
  • High sample throughput for routine quality control.
  • Excellent retention time reproducibility.
  • Lower operating pressure compared to sub-2 µm columns, eliminating the need for specialized pumps.
  • Robust performance for diverse ketone mixtures in pharmaceutical or environmental testing.

Future Trends and Opportunities


Emerging areas for development include:
  • Expansion of core–shell technology to a wider range of polar and non-polar analytes.
  • Integration with mass spectrometry for enhanced sensitivity and specificity.
  • Green chromatography approaches using reduced solvent volumes and alternative mobile phases.
  • Automation of sample preparation and direct online monitoring applications.

Conclusion


The Accucore C18 column delivers rapid, high-efficiency separation of ketones with excellent peak shape and reproducibility. Its compatibility with conventional HPLC hardware and significant reduction in analysis time make it an attractive choice for high-throughput laboratories.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF