LC-MS: Ultra-robust micro-flow LC-MS/MS for targeted high-throughput peptide quantification using the Vanquish Neo UHPLC system

Technical notes | 2021 | Thermo Fisher ScientificInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Proteomics
Manufacturer
Thermo Fisher Scientific

Significance of the Topic


Increasing throughput and robustness in peptide quantification is critical for large scale proteomic validation studies where nano flow LC MS methods face limitations in sample loading and reproducibility.

Objectives and Study Overview


This work adapts a previously published high throughput micro flow LC MS MS approach to the new Vanquish Neo UHPLC platform aiming at a robust method capable of quantifying standard peptides at 100 samples per day with consistent performance.

Methodology


HeLa protein digest spiked with heavy labeled PRTC peptides was prepared at 50 ng per microliter and 25 fmol peptide concentration. Samples were analyzed in direct injection micro flow mode with a 1.0 mm ID x 15 cm column, operating 100 microliters per minute for loading and washing and 50 microliters per minute during the gradient phase. Gradient conditions delivered 1 percent to 35 percent organic over 8.5 minutes with an overall cycle time of 14.4 minutes. Selected reaction monitoring transitions were monitored for twelve PRTC peptides. A sequence of 760 injections over 7.5 days assessed retention time, pressure and peak area stability.

Used Instrumentation


  • Vanquish Neo UHPLC system with Binary Pump N and Split Sampler NT
  • Acclaim PepMap 1.0 mm ID x 15 cm column
  • Thermo Scientific TSQ Altis triple quadrupole mass spectrometer with OptaMax NG HESI source
  • 50 micrometer ID nanoViper capillaries for reduced post column volume
  • Chromeleon 7.2.10 or Xcalibur with SII software for data acquisition

Main Results and Discussion


System pressure traces remained superimposable over the entire sequence with less than 3 bar variation. Retention time relative standard deviations were below 0.5 percent for all monitored peptides. Peak area reproducibility was below 10 percent RSD for ten peptides and below 15 percent for the lowest abundant targets. No performance decline was observed over 760 injections.

Benefits and Practical Applications


This micro flow LC MS MS protocol delivers high throughput, excellent reproducibility and robust operation suitable for large sample cohorts in translational proteomics and bioanalytical validation workflows.

Future Trends and Potential Applications


Further integration with automation and sample multiplexing could push throughput beyond 100 samples per day. Advances in column technology and mass spectrometer detection may enable deeper proteome coverage at micro flow rates and shorter cycle times.

Conclusion


The adapted micro flow method on the Vanquish Neo UHPLC system provides a robust, reproducible and high throughput solution for targeted peptide quantification in large scale studies.

Reference


  • Boychenko A et al, Technical Note 73208 Tailored high throughput low flow LC MS methods, Thermo Scientific 2019
  • Meding S, Boychenko A, Chromatography Today 2016 3 43-45
  • Bian Y et al, Robust reproducible quantitative proteomics by micro flow LC MS MS, Nat Commun 2020 11 157
  • Bian Y, Bayer FP et al, Robust Microflow LC MS MS for Proteome Analysis 38000 Runs Anal Chem 2021 93 8 3686-3690
  • European Medicines Agency Guideline on bioanalytical method validation EMEA CHMP EWP 192217 2009

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Quantitative targeted nano- and capillary- flow LC-MS peptide analysis using the Vanquish Neo UHPLC System coupled to a triple quadrupole mass spectrometer
LC-MS: Vanquish Neo UHPLC system-to-system reproducibility ensures consistent and reliable results in nanoLC-MS proteomics
Long-term stability and reproducibility of nano, capillary and micro-flow LC-MS separations
Long-term stability and reproducibility of nano-, capillary- and micro-flow LC-MS separations: the impact of hardware and separation column