Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-PAHs) in APCI mode with a single quadrupole mass spectrometer

Applications | 2018 | Thermo Fisher ScientificInstrumentation
LC/MS, LC/SQ
Industries
Environmental
Manufacturer
Thermo Fisher Scientific

Importance of the Topic


Polycyclic aromatic hydrocarbons (PAHs) and their oxygenated derivatives (oxy-PAHs) are widespread environmental contaminants with significant toxicity and carcinogenicity. Reliable detection and quantification of oxy-PAHs in water, soil, air, and food are critical for environmental monitoring, food safety, and human health risk assessment. While LC–MS with electrospray ionization (ESI) is common, polar oxy-PAHs often suffer from poor ionization and adduct formation. Atmospheric pressure chemical ionization (APCI) provides an alternative that can enhance sensitivity and simplify spectra for these challenging analytes.

Objectives and Overview of the Study


This study evaluates the performance of APCI versus ESI on a Thermo Scientific ISQ EM single quadrupole mass spectrometer for six standard oxy-PAHs. The goals are to:
  • Demonstrate APCI-MS detection of oxy-PAHs without derivatization.
  • Compare sensitivity, signal-to-noise (S/N), and adduct formation between APCI and ESI.
  • Establish a robust UHPLC separation for oxy-PAHs.

Methodology


Chromatographic Conditions
  • Column: Hypersil Green PAH, 150 × 2.1 mm, 3 µm.
  • Mobile phase: Water (A) and Methanol (B), flow rate 0.35 mL/min.
  • Gradient: 25% B to 100% B over 6 min, hold to 8 min, re-equilibrate to 25% B by 11 min.
  • Column temp: 40 °C; autosampler temp: 20 °C; injection: 1 µL.
  • UV detection at 254 nm confirmed baseline separation (resolution > 2).

Instrumentation Used


  • Thermo Scientific Vanquish Flex UHPLC system (Binary Pump F, Split Sampler HT, Column Compartment H, Variable Wavelength Detector F).
  • Thermo Scientific ISQ EM single quadrupole mass spectrometer with interchangeable APCI and ESI probes.
  • Chromeleon 7.2.9 CDS for data acquisition and processing.

Main Results and Discussion


  • APCI detected all six oxy-PAHs in positive and/or negative mode without derivatization; ESI full-scan missed one analyte (S/N ≈ 12).
  • Electron capture negative ionization in APCI enabled ionization of compounds lacking abstractable hydrogens (e.g., benzanthrone, 1,4-anthraquinone), which were not observed by negative ESI.
  • APCI delivered higher S/N ratios than ESI for most analytes, particularly in positive mode.
  • Fewer adducts formed in APCI simplified mass spectra and improved quantitation accuracy.
  • Dedicated PAH column achieved clear separation of target analytes and trace impurities.

Benefits and Practical Applications


  • Enhanced detection sensitivity and reliability for oxy-PAHs in environmental and food analysis.
  • Elimination of derivatization steps streamlines sample preparation and reduces analysis time.
  • Improved method robustness and reproducibility on a single quadrupole platform.
  • Broader applicability to polar combustion-derived contaminants in routine monitoring.

Future Trends and Potential Applications


  • Adoption of high-resolution MS with APCI for non-targeted screening of emerging oxy-PAHs.
  • Integration of automated sample preparation and UHPLC–MS workflows for higher throughput.
  • Development of regulatory-compliant quantitation methods for water and soil matrices.
  • Miniaturized or portable LC–APCI–MS systems for on-site environmental analysis.

Conclusion


APCI on a single quadrupole mass spectrometer significantly outperforms conventional ESI for the analysis of oxygenated PAHs. It provides superior S/N, reduced adduct formation, and the ability to ionize challenging analytes, offering a valuable tool for environmental, food safety, and industrial laboratories.

Reference


1. Lundstedt S., et al. Trends Anal. Chem. 2014, 57, 83–92.
2. Singh L., Agarwal T. Trends Food Sci. Technol. 2018, 79, 160–170.
3. Walgraeve C., et al. Anal. Bioanal. Chem. 2017, 409, 335–347.
4. Cochran R.E., Smoliakova I.P., Kubatova A. Int. J. Mass Spectrom. 2016, 397–398, 6–17.
5. Kumar S., Negi S., Maiti P. Environ. Sci. Pollut. Res. 2017, 24, 25810–25827.
6. Grosse S., Letzel T. J. Chrom. A 2007, 1139, 75–83.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Seeking More Productive Chromatography? See Productivity in a New Light
Thermo Scientific ISQ EC and ISQ EM single quadrupole mass spectrometers
Thermo Scientific ISQ EC and ISQ EM single quadrupole mass spectrometers
2018|Thermo Fisher Scientific|Brochures and specifications
Analysis of multiple matrices with a single calibration curve for polycyclic aromatic hydrocarbons (PAHs) with the ISQ 7610 GC-MS system following EPA Method 8270E
Sensitive and reproducible analysis of 16 polyaromatic hydrocarbons (PAHs) using gas chromatography – triple quadrupole mass spectrometry (GC-MS/MS)