Sample Preparation, Consumables, LC/MS, LC/MS/MS, LC/QQQ
IndustriesClinical Research
ManufacturerThermo Fisher Scientific
Importance of the Topic
Vitamin D status is routinely monitored as a critical indicator of bone health and calcium homeostasis in clinical settings. Measurement of circulating 25-hydroxyvitamin D2 and 25-hydroxyvitamin D3 provides a reliable assessment of total vitamin D status due to their higher blood concentrations and longer half life compared to the active hormone form.
Objectives and Study Overview
The aim of this work was to develop a rapid, sensitive and reproducible method for quantifying 25-hydroxyvitamin D2 and D3 in human plasma. This assay leverages a novel hydrophobic retention SPE product, SOLA HRP 30 mg, in conjunction with a UHPLC MS MS platform to improve sensitivity and throughput for clinical research applications.
Methodology and Sample Preparation
A small volume of plasma was protein precipitated with acetonitrile and fortified with an isotopically labeled internal standard. Samples were processed on SOLA HRP 30 mg SPE cartridges using a water methanol wash and methanol elution. Extracts were dried and reconstituted in a water acetonitrile mixture prior to analysis.
Used Instrumentation
- Vanquish Horizon UHPLC system with binary pump split sampler and column compartment
- Hypersil GOLD VANQUISH C18 UHPLC column, 1.9 micrometer, 2.1 by 50 millimeter
- TSQ Quantiva triple stage quadrupole mass spectrometer with HESI II source
- Chromeleon Chromatography Data System software, version 7.2.10
Key Results and Discussion
- Calibration was linear from 5 to 1000 nanograms per milliliter for D2 and from 2.5 to 1000 for D3 with correlation coefficients above 0.99
- Lower limits of quantification were 5.0 nanograms per milliliter for D2 and 2.5 for D3
- Intra day precision (CV) did not exceed 9.6 percent and accuracy bias was within ±8.2 percent at quality control levels
- Recovery at mid QC was approximately 100 percent for both analytes
- Matrix effects were minimal with internal standard normalization factors near unity
- No carryover was observed in blank injections following the highest calibration standard
Benefits and Practical Applications
- Enhanced sensitivity due to higher SPE bed mass and reduced breakthrough
- High throughput processing compatible with 96 well plate format
- Low sample and solvent consumption
- Robust method suitable for large scale clinical research and routine monitoring
Future Trends and Opportunities
- Extension to measurement of additional vitamin D metabolites
- Automation of sample preparation to increase throughput and reproducibility
- Integration into routine clinical laboratory workflows for population studies
- Investigation of lower quantification limits using larger sample volumes or enhanced SPE formats
Conclusion
The combination of SOLA HRP 30 mg SPE and UHPLC MS MS on a Vanquish TSQ Quantiva platform yields a fast, robust and sensitive assay for 25-hydroxyvitamin D2 and D3 in human plasma, offering excellent reproducibility and minimal matrix interference for clinical research.
Reference
- Holick MF, Binkley NC, Bischoff Ferrari HA, Gordon CM, Hanley DA, Heaney RP, Murad MH, Weaver CM. Evaluation, treatment, and prevention of vitamin D deficiency an Endocrine Society clinical practice guideline. J Clin Endocrinol Metab. 2011;96(7):1911–1930.
- Institute of Medicine Food and Nutrition Board. Dietary Reference Intakes for Calcium and Vitamin D. Washington DC National Academy Press; 2010.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.