LC/HRMS, LC/MS, LC/MS/MS, LC/Orbitrap
IndustriesClinical Research
ManufacturerThermo Fisher Scientific, RECIPE
Importance of Topic
Therapeutic drug monitoring of antidepressants supports optimized dosing and safety by accurately quantifying drug levels in plasma. High-resolution accurate mass (HRAM) LC-MS enables selective and sensitive analysis, increasing confidence in clinical research data.
Objectives and Study Overview
- Develop a single LC-HRAM MS method for quantifying 36 antidepressants in human plasma.
- Implement the method on a Thermo Scientific Orbitrap Exploris 120 MS coupled to a Vanquish Flex UHPLC.
- Validate analytical performance: linearity, accuracy, precision and lower limits of quantification (LLOQ).
Methodology
- Sample preparation by protein precipitation: 50 µL plasma with 100 µL acetonitrile containing deuterated internal standards.
- Chromatography on a Hypersil GOLD Phenyl column (2.1 × 100 mm, 1.9 µm) using a 0.5 mL/min gradient of 5 mM ammonium formate with 0.1% formic acid in water (A) and methanol (B).
- Mass spectrometry in full-scan positive mode (m/z 100–500) at 60,000 resolution using heated electrospray ionization (HESI).
- Data acquisition and quantification performed with Thermo Scientific TraceFinder 5.1 software.
Used Instrumentation
- Thermo Scientific Vanquish Flex Binary UHPLC system
- Thermo Scientific Hypersil GOLD Phenyl analytical column (2.1 × 100 mm, 1.9 µm)
- Thermo Scientific Orbitrap Exploris 120 mass spectrometer with HESI source
- TraceFinder 5.1 data processing software
Main Results and Discussion
- Linearity: calibration curves for all compounds showed R² > 0.993 with 1/x weighting.
- Accuracy of back-calculated calibrators ranged from 92.9 to 106.6% and precision < 7.3% across three levels.
- LLOQs extended below the lowest calibrator, meeting 80-120% accuracy and <20% CV for all analytes.
- Intra-assay precision and accuracy: 85.2–113.9% accuracy, CV < 6.5% (n=5, three days).
- Inter-assay precision and accuracy: 89.2–114.8% accuracy, CV < 6.4% (n=15).
Benefits and Practical Applications
- Rapid analysis: <7 minutes per sample throughput.
- Simple sample preparation with offline protein precipitation.
- High sensitivity and selectivity for comprehensive monitoring of multiple antidepressant classes.
- Supports clinical research and therapeutic drug monitoring workflows.
Future Trends and Potential Applications
- Expansion to include metabolites and additional psychotropic compounds.
- Integration with automated sample preparation and multiplexed assays.
- Application in real-time clinical decision support and personalized medicine.
- Adoption of AI-driven data interpretation to streamline HRAM MS workflows.
Conclusion
A robust LC-HRAM MS method on the Orbitrap Exploris 120 enables fast, accurate quantification of 36 antidepressants in human plasma. The validated workflow meets stringent performance criteria and offers a valuable tool for therapeutic drug monitoring in clinical research.