LC/HRMS, LC/MS, LC/MS/MS, LC/Orbitrap
IndustriesClinical Research
ManufacturerThermo Fisher Scientific, RECIPE
Significance of the Topic
Measurement of 25-hydroxyvitamin D2 and D3 informs vitamin D status impacting bone health, immune function, and chronic disease risk. High-resolution LC-MS overcomes limitations of protein binding bias and cross-reactivity in immunoassays and provides selective, sensitive quantification for clinical research.
Study Objectives and Overview
The aim was to develop and validate a rapid, robust LC-HRAM-MS method using a Thermo Scientific Orbitrap Exploris 120 to quantify 25-OHD2 and 25-OHD3 in human plasma. Method performance was assessed for linearity, sensitivity, accuracy, precision, and carryover.
Methodology and Sample Preparation
Plasma (50 µL) was spiked with a d6-25-OHD3 internal standard and precipitated with acetonitrile. After centrifugation the supernatant was directly injected.
Chromatographic separation used a Vanquish Flex UHPLC system with a Hypersil GOLD 2.1×50 mm column at 40 °C. Gradient elution with 5 mM ammonium formate/0.1% formic acid in water (A) and methanol (B) at 0.5 mL/min yielded a 3.5 min run time.
Mass spectrometry on Orbitrap Exploris 120 with an APCI positive ion source acquired full scan (100–500 m/z at 60 000 resolution) and targeted MS2 (30 000 resolution) with HCD at 40% collision energy.
Used Instrumentation
- Thermo Scientific Vanquish Flex UHPLC
- Hypersil GOLD 2.1×50 mm, 1.9 µm column
- Thermo Scientific Orbitrap Exploris 120 MS with APCI source
- TraceFinder 5.1 data processing software
Main Results and Discussion
The calibration curve used quadratic fitting with 1/x weighting; bias remained within ±10% across all calibrators. LLOQs were 1.97 µg/L (25-OHD2) and 4.52 µg/L (25-OHD3) for both acquisition modes. Accuracy bias ranged 1.9–5.8% (full scan) and 1.8–6.7% (MS2). Intra-assay precision (%CV) was ≤7.5% (full scan) and ≤8.4% (MS2); inter-assay precision was ≤5.0% and ≤5.9%, respectively. No significant carryover was observed.
Benefits and Practical Applications
- High selectivity and sensitivity enable reliable quantification of vitamin D metabolites.
- Rapid sample preparation and short run times support high throughput.
- Robust performance suits clinical research and biomarker studies.
Future Trends and Opportunities
Advances may include automation of sample preparation, multiplexed assays for additional metabolites, integration into high-throughput workflows, and application in large-scale epidemiological studies. Emerging high-resolution platforms and enhanced ionization techniques will further improve sensitivity and selectivity.
Conclusion
A streamlined LC-HRAM-MS method was validated for precise, accurate quantification of 25-OHD2 and 25-OHD3 in plasma. The technique offers robust performance suitable for clinical research quantifying vitamin D status.
Reference
- Mora JR, Iwata M, von Andrian UH. Vitamin effects on the immune system: vitamins A and D take centre stage. Nat Rev Immunol. 2008;8:685-698.
- Plum LA, DeLuca HF. Vitamin D, disease and therapeutic opportunities. Nat Rev Drug Discov. 2010;9:941–955.
- DeLuca HF. The vitamin D story: a collaborative effort of basic science and clinical medicine. FASEB J. 1988;2:224–236.
- Holick MF. Vitamin D Deficiency. N Engl J Med. 2007;357:266–281.
- Ohyama Y, Yamasaki T. Eight cytochrome P450s catalyze vitamin D metabolism. Front Biosci. 2004;9:3007–3018.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.