Cell Culture Optimization Using an Agilent Bio-Monolith Protein A Column and LC/MS

Applications | 2017 | Agilent TechnologiesInstrumentation
Consumables, HPLC, LC/MS, LC/MS/MS, LC columns, LC/QQQ
Industries
Pharma & Biopharma
Manufacturer
Agilent Technologies

Importance of Topic


Monoclonal antibodies represent a leading class of biotherapeutics, and their glycosylation profiles critically influence efficacy, stability, and immunogenicity. Matching biosimilar glycoforms to originator products is essential for regulatory approval and clinical performance.

Study Objectives and Overview


This work aimed to optimize Chinese hamster ovary (CHO) cell culture conditions for a trastuzumab biosimilar, focusing on achieving an originator-equivalent glycosylation pattern. The study varied concentrations of galactose, uridine, and manganese chloride and employed an Agilent Bio-Monolith Protein A column with LC/MS analysis to assess antibody titer and structural attributes.

Instrumentation


Bio-Monolith Protein A Analysis
  • Agilent 1100 Series Quaternary Pump, Autosampler, Diode Array Detector
  • Agilent 1200 Infinity Series Analytical-scale Fraction Collector
LC/MS Analysis
  • Agilent 1290 Infinity Binary LC (Binary Pump, Autosampler, Thermostat)
  • Agilent 6540 UHD Accurate-Mass Q-TOF with Jet Stream
Software
  • OpenLAB CDS ChemStation
  • MassHunter for instrument control and data analysis
  • BioConfirm for glycoform deconvolution


Main Results and Discussion


Protein A chromatograms showed distinct antibody peaks under varying feed conditions. Calibration with Herceptin demonstrated excellent linearity (0.02–2 mg/mL, R²=0.9999). High-resolution MS of reduced heavy chains identified six principal glycoforms (Man5, G0, G0F, G1F, G2F, G0F-GlcNAc). Originator batches exhibited lot-to-lot average G0F and G1F contents of 44.7±6.7% and 38.6±4.8%, respectively. Biosimilar cultures with increased galactose, uridine, and manganese showed elevated galactosylation (G1F:G0F ratio) up to 13.3% G2F at the highest feed. Antibody titers decreased from 0.47 mg/mL at 4× feed to 0.14 mg/mL at 24×.Optimal feed levels (4×–8×) balanced glycan profile alignment with acceptable yields.

Method Benefits and Practical Applications


This workflow integrates rapid Protein A capture with LC/MS for simultaneous quantification and glycoform profiling, streamlining clone selection and medium optimization. It ensures biosimilar quality attributes align with regulatory requirements and supports high-throughput bioprocess development.

Future Trends and Potential Applications


Future developments may include multi-attribute LC/MS assays combining intact mass, peptide mapping, and glycopeptide analysis for deeper characterization. Real-time online MS monitoring of glycosylation and automated feed strategies could further accelerate process development for novel antibodies and biosimilars.

Conclusion


The Agilent Bio-Monolith Protein A column coupled to LC/MS effectively guided CHO culture optimization, achieving trastuzumab glycosylation within originator specifications while maintaining viable antibody titers under intermediate feed conditions.

References


  1. K. Sandra, I. Vandenheede, P. Sandra. J. Chromatogr. A, 1335, 81 (2014).
  2. European Medicines Agency, infliximab biosimilars approval (2013).
  3. E. Dumont et al., Agilent Application Note 5991-5124EN (2014).
  4. Genentech information on trastuzumab HER2 mechanism.
  5. M.J. Gramer et al., Biotechnol. Bioeng. 108, 1591 (2011).

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Automated Switching between Peptide and Glycan Mapping with the Agilent 1290 Infinity II LC System
Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS
Prepping Biosimilars
Prepping Biosimilars
2014|Agilent Technologies|Brochures and specifications
Determination of Multiple Attributes of Monoclonal Antibodies