PPCP Analysis in Bottled, Tap and Surface Water using Time De-Coupled Chromatography

Posters | 2014 | Waters | HPLC SymposiumInstrumentation
HPLC
Industries
Environmental
Manufacturer
Waters

Importance of the Topic



Contamination of drinking and surface waters by pharmaceuticals and personal care products (PPCPs) poses risks to ecosystems and human health. Trace-level monitoring of these micropollutants requires highly sensitive, rapid and robust analytical workflows to support environmental surveillance and regulatory compliance.

Study Objectives and Overview



This work evaluates Time De-Coupled Chromatography (2D mode) coupled with LC-MS/MS for quantifying a broad suite of PPCPs in bottled, tap and surface water. Key aims are to minimize sample preparation time, reduce required sample volume and achieve sub-ppt detection limits.

Methodology and Instrumentation



Sample Preparation and Chromatography
  • Solid-phase extraction (SPE) using mixed sorbents (Oasis HLB, MAX, MCX) for simultaneous enrichment of acidic and basic PPCPs.
  • Automated AT-column dilution for direct injection of aqueous or organic extracts (up to 1 mL) without evaporation/reconstitution.
  • Time De-Coupled 2D chromatography enabling large-volume injections and fast transfer between trapping and analytical columns.
Instrumental Setup
  • Waters ACQUITY UPLC system with two binary pumps, 1000 µL injection loop and trap column.
  • BEH C18 and C8 columns for orthogonal separations.
  • Triple quadrupole MS operated in multiple reaction monitoring (MRM) mode with electrospray ionization (ES+).

Main Results and Discussion



The method delivered detection limits down to 1 ppt for multiple PPCPs in 15 mL water samples. Spike recoveries at 10 ppt ranged from 70 % to 115 %, with linearity (r2) above 0.99. Analysis time per sample was reduced from over one hour to under 15 minutes. Direct injection of up to 1 mL extract provided 10× enrichment, eliminating lengthy evaporation steps. Precision and accuracy were maintained below 10 % RSD across water matrices.

Benefits and Practical Applications


  • Trace-level detection (ppt) of diverse PPCPs.
  • Minimal sample preparation (< 15 min) and reduced manual labor (~80 % time savings).
  • Wide pH compatibility for acidic and basic analytes with mixed SPE sorbents.
  • High throughput enabling environmental monitoring and QA/QC in water utilities and research laboratories.
  • Robustness and reproducibility for routine analysis.

Future Trends and Applications



Further automation of SPE and on-line 2D coupling could enable field-deployable systems. Integration with high-resolution MS and data-driven quantification will broaden target lists. Green extraction solvents and miniaturized formats may reduce environmental footprint while extending the approach to other trace contaminants.

Conclusion



Time De-Coupled Chromatography combined with LC-MS/MS offers a rapid, sensitive and robust solution for ppt-level PPCP analysis in various water matrices. The streamlined workflow drastically cuts analysis time and sample volume, supporting high-throughput environmental monitoring and regulatory testing.

References


  • Mallet CR. PPCP Analysis in Bottled, Tap and Surface Water using Time De-Coupled Chromatography. Waters Corporation; 2014.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Analysis of Pharmaceuticals and Pesticides in Bottled, Tap, and Surface Water Using ACQUITY UPLC Systems with 2D-LC Technology
Multi-Residue Analysis of Pharmaceuticals and Personal Care Products (PPCPs) in Water Using the ACQUITY UPLC H-Class System and the Xevo TQD Tandem Mass Spectrometer
MULTI-DIMENSIONAL CHROMATOGRAPHY COMPENDIUM - TRAP AND ELUTE VS AT-COLUMN DILUTION