Analysis of Tricyclic Antidepressants in Human Urine Using UltraPerformance Convergence Chromatography (UPC2)

Applications | 2013 | WatersInstrumentation
LC/MS, LC/MS/MS, LC/QQQ, SFC
Industries
Clinical Research
Manufacturer
Waters

Significance of the Topic


Tricyclic antidepressants (TCAs) remain widely used due to their cost-effectiveness and clinical efficacy in cases where newer agents are inadequate. Accurate, high-throughput quantification of TCAs in biological fluids is essential for therapeutic drug monitoring, pharmacokinetic studies, and compliance testing. UltraPerformance Convergence Chromatography (UPC²) introduces a greener, faster alternative to traditional reversed-phase LC, leveraging supercritical CO₂ to reduce solvent use and improve selectivity.

Objectives and Study Overview


This work presents a proof-of-concept Application Note demonstrating a rapid UPC²–MS/MS method for simultaneous analysis of five TCAs in human urine. The goals were to minimize organic solvent consumption, simplify sample preparation through direct injection of SPE eluates, and achieve sub-nanogram per milliliter sensitivity in a 3-minute chromatographic cycle.

Methodology and Instrumentation


Sample preparation employed Oasis WCX 96-well µElution plates for mixed-mode weak cation exchange extraction of acidified urine, followed by direct injection of combined elution aliquots without evaporation.
  • Chromatography: Waters ACQUITY UPC² system with a BEH 2.1 × 50 mm, 1.7 µm column.
  • Mobile phases: CO₂ (A) and methanol + 0.2% NH₄OH (B); gradient from 2% to 40% B in 2 min, total run time 3 min.
  • Mass spectrometry: Xevo TQ-S triple quadrupole, positive ESI, optimized cone voltages and collision energies for each analyte.

Key Results and Discussion


The optimized method provided rapid separation with peak widths <2 s at base and low system pressures (<4200 psi). SPE recoveries ranged from 92% to 104% (RSD 3%–6%). Systematic screening identified the BEH column and NH₄OH modifier as optimal for sensitivity and peak shape. Calibration curves over 0.1–10 ng/mL were linear (R² > 0.998) with average accuracy within ±8%. LLOQs of 0.1 ng/mL met FDA bioanalytical criteria, with signal-to-noise ratios from 66:1 to 590:1.
  • High throughput: 3 min cycle time without carryover.
  • Excellent sensitivity: sub-ng/mL quantification in urine.
  • Robust performance: reproducible retention times and peak areas.

Benefits and Practical Applications


  • Green chemistry: supercritical CO₂ reduces organic solvent consumption and waste.
  • Workflow efficiency: direct injection of SPE eluates eliminates evaporation and reconstitution steps.
  • Wide applicability: suitable for therapeutic drug monitoring, pharmacokinetics, and toxicology screening.

Future Trends and Opportunities


UPC² technology is expanding its role in bioanalysis, offering orthogonal selectivity to reversed-phase LC and compatibility with diverse sample matrices. Potential advancements include:
  • Implementation of buffered modifiers (e.g., ammonium formate) to improve peak shapes for polar compounds.
  • Integration with fully automated sample preparation for large-scale studies.
  • Extension to other drug classes and metabolite panels for comprehensive screening.

Conclusion


The UPC²–MS/MS method delivers a fast, accurate, and environmentally sustainable approach for quantifying tricyclic antidepressants in human urine. The combination of efficient mixed-mode SPE, optimized UPC² separation, and sensitive MS detection fulfills stringent bioanalytical requirements while enhancing throughput and reducing solvent usage.

References


  1. Baker GB, Coutts RT, Holt A. Derivatization with acetic anhydride: Applications to the analysis of biogenic amines and psychiatric drugs by gas chromatography and mass spectrometry. Journal of Pharmacological and Toxicological Methods. 1994;31(3):141-148.
  2. Cantú M, Toso D, Lacerda C, Lanças F, Carrilho E, Queiroz M. Optimization of solid-phase microextraction procedures for the determination of tricyclic antidepressants and anticonvulsants in plasma samples by liquid chromatography. Analytical and Bioanalytical Chemistry. 2006;386(2):256-263.
  3. Santos-Neto AJ, Bergquist L, Lanças FM, Sjöberg PJ. Simultaneous analysis of five antidepressant drugs using direct injection of biofluids in a capillary restricted-access media-liquid chromatography–tandem mass spectrometry system. Journal of Chromatography A. 2008;1189(1-2):514-522.
  4. Chambers E, Wagrowski-Diehl DM, Lu Z, Mazzeo JR. Systematic and comprehensive strategy for reducing matrix effects in LC/MS/MS analyses. Journal of Chromatography B. 2007;852(1-2):22-34.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Ultrafast Analysis of a Tricyclic Antidepressant Drug Panel in Human Serum by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System
Quantification of 15 tricyclic antidepressants in human plasma by liquid chromatography-tandem mass spectrometry for clinical research
Quantification of 15 tricyclic antidepressants in human plasma or serum by LC-HRAM-MS for clinical research
Quantification of 15 tricyclic antidepressants in human plasma by LC-HRAM-MS using the Orbitrap Exploris 120 mass spectrometer for clinical research