LC/TOF, LC/HRMS, LC/MS, DART
IndustriesPharma & Biopharma
ManufacturerWaters
Significance of the Topic
Ambient desorption ionization methods like ASAP enable rapid, minimal-prep detection of synthetic PDE-5 inhibitors in counterfeit and adulterated products, addressing public health threats and high-throughput analytical needs.
Objectives and Overview of the Study
This study applied the Waters Atmospheric Solids Analysis Probe (ASAP) coupled with high-resolution mass spectrometry to rapidly detect and identify synthetic phosphodiesterase type-5 inhibitors in counterfeit erectile dysfunction tablets and adulterated herbal supplements.
Methodology and Instrumentation
- Sample selection: Counterfeit brand and generic ED tablets and five herbal dietary supplements purchased online.
- ASAP-MS analysis: Samples applied to glass capillaries and introduced into a heated gas stream (100–450 °C) for vaporization; corona discharge ionization in APCI mode.
- Instrument: Waters LCT Premier XE high-resolution TOF-MS operating in combined ESCi (ESI + APCI) mode; key parameters: corona current 5 µA, desolvation gas 500 L/hr, acquisition range 100–1000 amu.
Main Results and Discussion
- Counterfeit tablets: ASAP-MS identified correct API in genuine tablets (sildenafil, vardenafil, tadalafil) and detected wrong or absent APIs in imitations within 20 minutes without chromatography.
- Herbal supplements: All five HDS samples contained undeclared sildenafil, tadalafil, or analogues such as thiohomosildenafil, confirmed by accurate mass and elemental composition.
- Comparison with UPLC-MS/MS: ASAP findings were consistent with detailed chromatographic analyses, validating the rapid screening approach.
Benefits and Practical Applications
- Minimal or no sample preparation reduces analysis time and bottlenecks associated with LC/GC separations.
- Rapid yes/no screening capability guides decisions on further chromatographic characterization.
- High-resolution mass measurement and elemental composition facilitate confident compound identification in counterfeit and adulterated products.
- Enhanced laboratory throughput supports regulatory and pharmaceutical quality control efforts.
Future Trends and Opportunities
- Integration with automated sampling systems for large-scale screening of pharmaceuticals and supplements.
- Expansion to other classes of illicit or adulterant compounds across food, cosmetics, and forensic matrices.
- Coupling with ion mobility or advanced fragmentation techniques to resolve isomers and novel analogues.
- Development of portable ASAP-MS platforms for in-field screening by customs and regulatory agencies.
Conclusion
ASAP coupled with high-resolution MS offers a robust, high-throughput screening tool for the rapid detection of synthetic PDE-5 inhibitors in counterfeit and adulterated products, significantly reducing time and resources compared to traditional chromatography-based methods.
References
- Takats Z, Wiseman JM, Gologan B, Cooks RG. Science. 2004;306:471.
- Cody R, Laramee J, Durst H. Anal Chem. 2005;77:2297.
- McEwen CN, McKay RG, Larsen BS. Anal Chem. 2005;77:7826.
- Pharmaceutical Counterfeiting. Analyst. 2005;130:271–279.
- Rudolf PM, Berinstein IBG. N Engl J Med. 2004;350(14):1384.
- Liang Q, Qu J, Luo G, Wang Y. Pharm Biomed Anal. 2006;40:305–311.
- Reepmeyer JC, Woodruff JT. J Chrom A. 2006;1125:67–75.
- Zou P, Sze-Yin Oh S, Hou P, Low MY, Koh HL. J Pharm Biomed Anal. 2008;48:1070–1075.
- Venhuis BJ, Blok-Tip L, de Kaste D. Forensic Sci Int. 2008;177:e25–e27.
- Singh S, Prasad B, Savaliya AA, Shah RP, Gohil VM, Kaur A. Trends Anal Chem. 2009;28:1.
- Reepmeyer JC, Andre d’Avignon D. J Pharm Biomed Anal. 2009;49:145–150.
- Waters Corporation Application Note 720003185en AG-PDF. 2009.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.