Consumables, LC/TOF, LC/HRMS, LC/MS, LC/MS/MS
IndustriesClinical Research
ManufacturerWaters
Importance of the Topic
Terminal sialic acids on N-glycans critically influence the stability, serum half-life, and immunogenic profile of glycoprotein therapeutics. Accurate profiling of sialylated glycan species is essential in biopharmaceutical development, quality control, and cell-culture monitoring.
Study Objectives and Overview
This application note describes a systematic investigation to enhance hydrophilic interaction chromatography (HILIC)–fluorescence detection (FLR)–mass spectrometry (MS) analysis of RapiFluor-MS labeled sialylated N-glycans. Key aims were to optimize MS source parameters, refine HILIC separation using high ionic strength mobile phases on a wide-pore amide bonded phase, and integrate neuraminidase treatments for linkage-specific sialic acid profiling.
Methodology and Instrumentation
Samples: RapiFluor-MS Sialylated Glycan Performance Test Standard (bovine fetuin N-glycans)
HILIC Separation:
- ACQUITY UPLC Glycoprotein BEH Amide Column, 300 Å, 1.7 µm, 2.1×150 mm
- Mobile phases: high-strength ammonium formate (50–200 mM, pH 4.4) in water, acetonitrile as organic modifier
- Gradient optimized to balance retention and resolution for up to tetrasialylated species
Detection:
- FLR: Ex 265 nm/Em 425 nm
- MS: Xevo G2-XS QTof, positive-ion ESI, m/z 700–2000 (narrowed to 715–2000 for high salt buffers)
- Software: MassLynx 4.1, UNIFI 1.8, LockSpray calibration
Neuraminidase Digestions:
- α2-3 Neuraminidase S for selective removal of α2-3 sialic acids
- α2-3,6,8,9 Neuraminidase A for complete sialic acid release
Key Results and Discussion
MS Source Parameter Optimization:
- Sampling cone voltage: 75 V yielded highest protonated tri-charged ion ([A3G3S3+3H]³⁺) while minimizing in-source fragmentation
- Desolvation temperature: 500 °C balanced signal intensity and fragmentation control
- Capillary voltage (2200 V), source temperature (120 °C), and desolvation gas flow (600 L/Hr) were confirmed as optimal
HILIC Resolution Enhancement:
- Increasing ammonium formate from 50 to 200 mM extended retention times and sharpened peaks for highly sialylated glycans
- Baseline resolution achieved for isomeric tri- and tetrasialylated species (e.g., A3G3S3 variants)
- High-salt mobile phases introduced low-mass background; narrowing MS acquisition window restored sensitivity
Sialic Acid Linkage Profiling:
- Native profile: predominantly tri- (72 %) and tetrasialylated (17 %) triantennary glycans plus disialylated biantennaries (11 %)
- After α2-3 digestion: disappearance of tetrasialylated peaks, indicating all α2-3 linkages in S4 species; quantitative ratios of α2-3 vs. α2-6 in di- and tri-sialylated glycans were inferred
- After broad-specificity digestion: only asialo bi- and triantennary cores remained, confirming complete sialic acid removal
Benefits and Practical Applications
- Enhanced MS sensitivity for sialylated glycan species through fine-tuned ionization settings
- Improved chromatographic resolution supports accurate quantification of isomeric glycans
- Rapid, fluorescence-MS compatible workflow suitable for routine glycan profiling in QC labs
- Neuraminidase-based linkage assay provides fast, cursory insight into sialic acid linkage distribution without extensive MS/MS
Future Trends and Applications
Advances could include integration with glycan retention libraries for automated identification, deployment of ion mobility for enhanced isomer separation, and coupling with high-throughput automation to accelerate glycan analysis in bioprocess monitoring and personalized medicine.
Conclusion
This work demonstrates a robust HILIC-FLR-MS platform for high-fidelity analysis of sialylated N-glycans. Optimized MS source parameters and high-strength mobile phases on a wide-pore amide column yield superior sensitivity and resolution, while neuraminidase treatments enable simple linkage interrogation. Together, these methods support comprehensive glycan characterization in therapeutic development and quality control.
References
- Varki A. Trends Mol Med. 2008;14(8):351–360.
- Ashwell G, Harford J. Annu Rev Biochem. 1982;51:531–563.
- Lauber MA et al. Anal Chem. 2015;87(10):5401–5409.
- Harvey DJ et al. Proteomics. 2009;9(15):3796–3801.
- Ahn J et al. J Chromatogr B. 2010;878(3):403–408.