HOW LOW CAN WE GO? ANALYSIS OF ALDOSTERONE USING A HIGHLY ANALYTICALLY SENSITIVE TANDEM QUADRUPOLE MASS SPECTROMETER FOR CLINICAL RESEARCH

Posters | 2016 | WatersInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Clinical Research
Manufacturer
Waters

Importance of the Topic


Aldosterone is a key mineralocorticoid hormone responsible for maintaining sodium/potassium balance and blood pressure. Accurate measurement at low concentrations is critical in clinical research on aldosterone synthase (CYP11B2) inhibitors. Traditional immunoassays can suffer cross-reactivity with structurally related steroids, whereas LC-MS/MS offers greater selectivity and sensitivity.

Study Objectives and Overview


This study compared the performance of a conventional Xevo TQ-S tandem quadrupole mass spectrometer and the new Xevo TQ-XS system, both coupled to UPLC, for quantifying aldosterone in human plasma down to 2 pg/mL using 200 µL sample volume. It also evaluated the novel UniSpray ionization against electrospray ionization (ESI) for enhancing analytical sensitivity.

Methodology


Sample Preparation:
  • Automated on a Tecan Freedom EVO 100 liquid handler.
  • Protein precipitation with zinc sulfate in methanol and 0.05 % phosphoric acid.
  • SPE cleanup using Waters Oasis MAX µElution plates.
Chromatography:
  • Waters ACQUITY UPLC I-Class system.
  • CORTECS UPLC C18 column (2.1×100 mm).
  • Water/methanol gradient at 0.3 mL/min.

Used Instrumentation


  • Waters Xevo TQ-S and Xevo TQ-XS tandem quadrupole MS.
  • UniSpray source with grounded probe and Coanda-effect gas flow.
  • Waters ACQUITY UPLC I-Class.
  • Tecan Freedom EVO 100.

Main Results and Discussion


Analytical Sensitivity and Linearity:
  • LLOQ at 1 pg/mL (2.8 pmol/L) with S/N ≥10 on Xevo TQ-XS (350 fg on column).
  • Over 5-fold S/N improvement and over 6-fold peak area increase versus Xevo TQ-S.
  • Calibration range 15–1500 pg/mL (42–4161 pmol/L), r² > 0.9999.
Precision and Accuracy:
  • QC levels at 21, 42, 174 and 669 pg/mL showed total precision ≤7.9 % RSD on Xevo TQ-XS versus ≤8.2 % on Xevo TQ-S.
System Comparison:
  • Deming regression between systems: y = 0.98x – 5.90, no significant bias (p > 0.05).
  • Altman-Bland mean bias of –4.2 %.
  • Linear correlation r = 0.999.
UniSpray Evaluation:
  • Provided >3× increase in S/N at 2 pg/mL versus ESI.
  • Peak area reproducibility <20 % RSD at 60 fg on column.

Benefits and Practical Applications


  • Enables reliable low-level aldosterone quantification for pharmacokinetic and pharmacodynamic studies.
  • Reduces sample volume requirements and increases confidence in data at trace concentrations.
  • UniSpray may lower maintenance downtime and extend analyte coverage.

Future Trends and Applications


  • Broader validation of UniSpray for diverse steroid panels.
  • Integration of high-throughput workflows for clinical bioanalysis.
  • Application to additional low-abundance hormones and metabolites.
  • Advancements in automation and miniaturized sample prep.

Conclusion


The Waters Xevo TQ-XS coupled to UPLC demonstrates superior sensitivity, precision and linearity for plasma aldosterone analysis compared to the Xevo TQ-S. UniSpray ionization shows promising gains in signal and robustness, offering a powerful platform for clinical research requiring low-level steroid quantification.

References


  1. Häggström M, Richfield D. Diagram of the pathways of human steroidogenesis. Wikiversity Journal of Medicine. 2014;1(1).

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Highly Analytically-sensitive Analysis of Aldosterone in Plasma on the Xevo TQ-XS for Clinical Research
Ultra-Low Level Analysis of Aldosterone in Plasma Using  the Xevo TQ-XS for Clinical Research
UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research
Analysis of Corticosteroids and Androgens in Serum  for Clinical Research