Significance of the topic
Accurate detection and quantification of N-nitrosodimethylamine (NDMA) in drinking water is critical due to its classification as a potent carcinogen at concentrations as low as 10 ng/L. Traditional GC-MS approaches often face challenges in balancing sensitivity and selectivity, leading to potential interferences and false positives when analyzing complex water matrices.
Study objectives and overview
This work presents a single-injection method using high-resolution accurate-mass GC Orbitrap technology for:
- Targeted quantification of NDMA at sub-nanogram per liter levels.
- Comprehensive non-targeted screening of additional contaminants in drinking water.
Methodology and instrumentation
Instrumentation and key settings:
- GC-MS system: Thermo Scientific Exactive GC Orbitrap with TRACE 1310 GC, TriPlus RSH autosampler and TraceGOLD TG-1701MS column (30 m × 0.25 mm × 0.25 µm).
- Ionization: Electron ionization at 70 eV; full-scan mass range m/z 50–650; resolving power 60 000 FWHM (m/z 200); mass accuracy <0.5 ppm using PFTBA calibration.
- Sample preparation: Solid phase extraction (1000× enrichment) of tap water spiked at 0.96–9.6 ng/L NDMA; d6-NDMA surrogate and d14-NDPA injection standard.
- Calibration: Solvent standards from 0.1 to 100 µg/L NDMA with deuterated internal standards; data processed with Thermo Scientific TraceFinder for both quantitation and automated deconvolution-based compound identification.
Main results and discussion
- Chromatography: 11-minute GC run time; clear peak shape for NDMA at m/z 74.04747; resolution >100 000 at m/z 74 resolved potential isobaric interferences.
- Sensitivity: Instrument detection limit (IDL) of 0.09 µg/L (0.09 ng/L equivalent post-SPE).
- Repeatability: <4.5 % RSD for d6-NDMA (n=35), ~5 % RSD for d14-NDPA (n=14).
- Linearity: Excellent linear response (R2 >0.999) over 0.1–50 µg/L; %RSD of response factors <9 % for NDMA and <6 % for the d6-NDMA surrogate.
- Mass accuracy: Consistently <1 ppm deviation for the NDMA molecular ion in standards and matrix samples.
- Quantification in water: NDMA recoveries of 88–111 % at 0.96–9.6 ng/L; surrogate recoveries within QC limits (70–130 %).
- Non-targeted screening: Identification of 220 additional compounds absent in procedural blank, predominantly halogenated by-products, pharmaceuticals and monoterpenes. Examples include chloroiodomethane and tetrachloroethylene with high library scores (>890) and mass accuracy <0.5 ppm.
Benefits and practical applications
This GC Orbitrap method offers rapid, robust analysis combining trace-level NDMA quantification and broad non-targeted screening in a single injection. High mass resolution and accuracy reduce false positives, making it well suited for routine environmental monitoring, regulatory compliance and QA/QC in water laboratories.
Future trends and applications
- Extension to other nitrosamines and emerging organic contaminants.
- Integration with larger accurate-mass spectral libraries and advanced data workflows for automated identification.
- Application to diverse water matrices such as wastewater and industrial effluents.
- Development of on-line SPE-GC Orbitrap systems for fully automated monitoring.
Conclusion
The Exactive GC Orbitrap platform achieves sub-ng/L detection and quantification of NDMA with outstanding repeatability, linearity and mass accuracy. Its full-scan capability enables simultaneous targeted and non-targeted workflows, providing a powerful tool for comprehensive drinking water analysis.
References
- Mitch WA, Sharp JO, et al. N-nitrosodimethylamine (NDMA) as a drinking water contaminant: A review. Environmental Engineering Science 20(5):389–404 (2003).
- Sedlak DL, Deeb RA, et al. Sources and fate of nitrosodimethylamine and its precursors in municipal wastewater treatment plants. Water Environment Research 77(1):32–39 (2005).
- U.S. Environmental Protection Agency. Method 521: Determination of Nitrosamines in Drinking Water by SPE and GC/MS with Large Volume Injection and CI-MS/MS (2004).
- Thermo Fisher Scientific. Discovery of Emerging Disinfection By-Products in Water Using GC-Orbitrap MS. Application Note 10490 (2015).
- U.S. EPA. Basic Information about Tetrachloroethylene in Drinking Water (2014).
- California Department of Health Services. Notification Levels for NDMA in Drinking Water (2002).
- Health Canada. Guidelines for Canadian Drinking Water Quality: N-Nitrosodimethylamine (2011).
- Krauss M, Hollender J. Analysis of nitrosamines in wastewater: performance of a hybrid linear ion trap/orbitrap MS. Analytical Chemistry 80:834–842 (2008).