Hydrogen as a GC carrier gas for PCDD/F - analysis with Agilent 7010 MS-MS systems: Implementation and Evaluation

Presentations | 2020 | Agilent TechnologiesInstrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Industries
Environmental
Manufacturer
Agilent Technologies

Significance of the Topic


Analysis of polychlorinated dibenzo-p-dioxins and furans (PCDD/Fs) is critical for food and feed safety due to their toxicity at extremely low levels.
Traditional methods rely on helium as a GC carrier gas, but rising costs and supply concerns motivate exploration of alternatives.
Using hydrogen can reduce analysis time and operational costs while maintaining regulatory compliance.

Study Objectives and Overview


This work evaluates hydrogen as carrier gas for PCDD/F analysis on Agilent 7010 triple quadrupole MS/MS systems.
The project covers:
  • Comparison of detector performance: Agilent 7000C vs 7010A under helium and hydrogen
  • Development of a final GC method incorporating a shortened capillary column
  • Validation on real food and feed samples
  • Long-term stability and safety assessment

Methodology and Instrumentation


Sample preparation followed isotope dilution with 13C-labeled standards for 17 target congeners.
Gas chromatography used a VF-Xms fused-silica column (initially 60 m×0.25 mm i.d.×0.25 µm, later 40 m×0.18 mm i.d.×0.18 µm) under helium or hydrogen.
Detection employed Agilent 7000C and the new 7010A ion source for MS/MS, monitoring characteristic transitions.
Instrument stabilization after switching to hydrogen required up to five days for baseline and signal stability.
Safety measures included installing hydrogen sensors, proper venting and controlled source purging.

Main Results and Discussion


Detector qualification showed the 7010A with helium achieved an order of magnitude lower iLOQ for 2,3,7,8-TCDD compared to 7000C.
Under hydrogen, the 7010A maintained acceptable sensitivity with roughly half the iLOQ loss versus helium.
Chromatographic run time decreased from 53 to 34.5 min, improving daily sample throughput by 52 %.
Calibration linearity and response factors were stable over the range 0.002–10 pg/µL, with RSDs below 13 % for critical congeners.
Correlation of PCDD/F levels in 167 real samples between helium and hydrogen methods showed slopes near unity (0.96–1.08) and R2 ≥ 0.94.
Long-term operation over five years and thousands of injections confirmed robustness, with routine iLOQ for 2,3,7,8-TCDD at 0.023 pg on column under hydrogen.
Caveats included minor dechlorination in injector and ion source under hydrogen, potentially affecting pattern recognition; impact on TEQ was negligible but requires monitoring near regulatory limits.

Benefits and Practical Applications


Using hydrogen as carrier gas delivers faster separations, increased laboratory productivity and significant cost savings due to lower gas expense.
The Agilent 7010A ion source compensates sensitivity loss, enabling regulatory-compliant LOQs for PCDD/Fs in food and feed.
The method supports routine monitoring and QA/QC in food safety laboratories facing helium shortages.

Future Trends and Potential Applications


Further refinement of hydrogen-based GC-MS methods may extend to other persistent organic pollutants and emerging contaminants.
Integration with automated sample preparation and data processing pipelines can enhance throughput and traceability.
Advances in ion source design and AI-driven spectral deconvolution may further mitigate hydrogen-related dechlorination effects.
Adoption of renewable hydrogen sources could improve sustainability in analytical laboratories.

Conclusion


The study demonstrates that hydrogen is a viable GC carrier gas for PCDD/F analysis on Agilent 7010 MS/MS systems.
Method performance meets regulatory requirements with faster run times and lower costs, provided adequate stabilization and safety controls are implemented.
Minor analytical biases from dechlorination are manageable and do not compromise routine compliance.

Reference


Commission Regulation (EU) 2017/644 and 2017/771 on methods for PCDD/F analysis in food and feed; Frank Neugebauer, Eurofins GfA Lab Service Live Webcast 15 October 2020

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Hydrogen as Carrier Gas for the Analysis of Dibenzo-p-dioxins and Dibenzofurans (PCDD/Fs) in Food and Feed Samples Using an Agilent 7010 Series Triple Quadrupole GC/MS
Low level quantification of PCDD/Fs in animal feed using the Thermo Scientific TSQ 9000 triple quadrupole GC-MS/MS system with AEI source
DIOXIN: Low level quantification of PCDD/Fs in animal feedstuffs using the Thermo Scientific TSQ 9000 GC-MS/MS system with AEI source
Low level quanti fication of PCDD/Fs in animal feedstuffs using the Thermo Scientific TSQ 9000 GC-MS/MS system with AEI source