Ultrafast Analysis of Synthetic Colorants

Applications | 2011 | ShimadzuInstrumentation
HPLC
Industries
Food & Agriculture
Manufacturer
Shimadzu

Significance of the Topic


The analysis of synthetic food colorants is critical for ensuring product compliance, consumer safety, and regulatory adherence. Rapid, reliable methods support high-throughput screening in food quality control laboratories and reduce operational costs.

Objectives and Study Overview


This application note illustrates an ultrafast gradient HPLC method for simultaneous separation of 12 commonly used synthetic food dyes. By leveraging a core–shell C18 column and advanced gradient delivery, the study aimed to shorten run times without sacrificing chromatographic performance.

Methodology and Instrumentation


The method uses Shimadzu’s Nexera UHPLC system equipped with a precision gradient mixer to deliver stable, high‐efficiency elution. A Phenomenex Kinetex C18 core–shell column (50 mm × 3.0 mm I.D., 2.6 µm) was employed to allow elevated linear velocities. Samples of 50 mg/L each colorant were injected in 1 µL volume. Detection was carried out using a UV‐VIS detector (SPD‐M20A) with maximum absorbance monitoring between 400–700 nm.
  • System: Shimadzu Nexera UHPLC
  • Column: Phenomenex Kinetex C18, 2.6 µm core–shell (50 × 3.0 mm)
  • Mobile phase A: 10 mmol/L ammonium acetate; B: A/acetonitrile (1:1)
  • Gradient: 5 % B to 100 % B in 0–1.1 min, hold to 1.6 min
  • Flow rate: 2.5 mL/min (20 µL mixer)
  • Column temperature: 40 °C
  • Detection: UV‐VIS (400–700 nm), semi‐micro flow cell
  • Operating pressure: ~53 MPa

Main Results and Discussion


The optimized UHPLC protocol achieved baseline separation of all 12 dyes within a 1-minute runtime—over 20 times faster than a conventional 20-minute gradient on a 150 × 4.6 mm, 4.6 µm column. Repeatability tests (n=6) showed retention time %RSD below 0.11 % and peak area %RSD below 0.29 % for all analytes, confirming robust performance.

Benefits and Practical Applications


  • Throughput enhancement: Up to 20-fold reduction in analysis time accelerates sample turnaround.
  • High precision: Excellent repeatability supports reliable quantitative analysis.
  • Resource efficiency: Lower solvent consumption and extended column lifetime reduce operational costs.
  • Quality control: Rapid screening improves detection of colorant adulteration or compliance issues.

Future Trends and Applications


Continued advances in core–shell particle design, mixer miniaturization, and UHPLC system pressure capabilities will push analysis times below one minute for complex mixtures. Integration with automated sample handling and MS detection can expand applicability to trace-level quantitation and multi-residue screening.

Conclusion


This ultrafast UHPLC approach offers a compelling solution for high-throughput, precise analysis of synthetic food colorants, meeting the increasing demands of modern food safety and quality control laboratories.