Quantitative Analysis of -Lactam Antibiotics in Human Plasma by High Sensitivity LC/MS/MS Method

Posters | 2017 | ShimadzuInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Clinical Research
Manufacturer
Shimadzu

Importance of the Topic


The accurate quantitation of β-lactam antibiotics in human plasma is essential for therapeutic drug monitoring, optimizing dosage, minimizing resistance development and improving patient outcomes.

Study Objectives and Overview


This work describes a high-sensitivity LC/MS/MS method for the simultaneous quantitation of five β-lactam antibiotics – meropenem, tazobactam, piperacillin, cefepime and ceftazidime – in human plasma. The goal was to develop a fast, robust assay with minimal sample volume, simple preparation and reliable performance metrics.

Methodology and Instrumentation


A protein precipitation protocol was applied by adding acetonitrile:methanol (1:1) to plasma, followed by vortexing, centrifugation and filtration. A 2 μL aliquot of supernatant was injected onto a Kinetex C18 column under a 5.5-minute gradient. Detection employed a Shimadzu LCMS-8060 triple quadrupole with heated ESI in positive MRM mode. Stable isotope-labeled analogs served as internal standards for each analyte (MER-d6 for tazobactam).

Instrumental Details


  • LC: Kinetex 1.7 μm C18, 100×2.1 mm, 0.5 mL/min, 40 °C
  • Mobile Phases: 0.1% formic acid in water (A), 0.1% formic acid in acetonitrile (B)
  • Gradient: 5% B (0–0.2 min) → 90% B (3.5–4.0 min) → 5% B (4.1–5.5 min)
  • MS: MRM positive mode, block 400 °C, nebulizing gas N2 3.0 L/min, drying gas N2 5.0 L/min

Main Results and Discussion


Calibration curves (20–4000 ng/mL) showed excellent linearity (R2 > 0.997). Accuracy ranged from 87% to 109% across concentrations, with repeatability (%RSD) generally below 6%. Recoveries exceeded 86% for all analytes. Initial matrix effects were notable for ceftazidime (33–40%) and tazobactam (128–149%) but improved (62–85% and 95–109%, respectively) after 2.5-fold dilution.

Benefits and Practical Applications


  • Rapid analysis with total run time under 6 minutes
  • High sensitivity (LOQs 0.7–6.0 ng/mL) with 2 μL injection volume
  • Simple protein crash sample prep supports high throughput
  • Isotope dilution ensures specificity and robust quantitation

Future Trends and Opportunities


Emerging advances may include ultra-high-pressure LC for further speed, HRMS for broader screening, microfluidic sample handling to reduce volume, and automation to integrate direct sampling and data processing in clinical settings.

Conclusion


A fast, sensitive LC/MS/MS assay was established for five β-lactam antibiotics in plasma, demonstrating linearity, accuracy, precision, and robustness. The method supports therapeutic drug monitoring with minimal sample and system maintenance burdens.

References


  1. Shaikh S., Fatima J., Shakil S., Rizvi S.M.D. 2015;22(1):90–101.
  2. Kamal M.A. Saudi Journal of Biological Sciences; Grebe S.K., Singh R.J. Clinical Biochemistry Reviews. 2011;32(1):5–31.
  3. Bhandarkar D. et al. ASMS 2015 Poster TP 280.
  4. Colin P., De Bock L., Tjollyn H., Boussery K., Van Bocxlaer J. Talanta. 2013;103:285–293.
  5. Barco S., Bandettini R., Maffia A., Tripodi G., Castagnola E., Giuliani G. Journal of Chemotherapy. 2015;27(6):343–347.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Shimadzu Solutions for Application Notebook - Clinical Research - Application Notebook
VITATOX: Therapeutic Drug Monitoring (TDM) of Antibiotics by LC MS/MS
Fully automated quantification of Meropenem, Tazobactam, Piperacillin and Dexamethasone in plasma
A Novel Platform of On-line Sample Pre-treatment and LC/MS/MS Analysis for Screening and Quantitation of Illicit Drugs in Urine