Sample Preparation, Consumables
IndustriesPharma & Biopharma
ManufacturerWaters
Significance of the Topic
Glycosylation profiling is a critical quality attribute in biopharmaceutical development, influencing drug safety, efficacy and manufacturing consistency. Rapid and reliable methods for released N-glycan labeling are essential to meet throughput demands in research and quality control laboratories.
Objectives and Study Overview
This study describes the automation of Waters GlycoWorks RapiFluor-MS N-glycan labeling workflows on the Andrew+ pipetting robot. Two protocols were implemented for medium-throughput (24 samples) and high-throughput (48 samples) analyses. Performance was compared to manual preparations, with targets of ≤25 percent deviation in total peak area and ≤5 percent in relative area.
Methodology and Instrumentation
The automated workflows employ the GlycoWorks RapiFluor-MS 24- and 48-Sample Kits for rapid fluorescent and mass-spectrometric detection of released N-glycans. Sample processing steps—denaturation, enzymatic release, labeling, HILIC cleanup and elution—were programmed in OneLab software and executed on the Andrew+ platform. The deck layout included multiple Tip Insertion System modules for different pipette volumes, a storage plate for reagent reservoirs, a vacuum-based microelution plate for cleanup, a Peltier-cooled PCR plate for reactions and microtube modules for enzymes and labeling reagents.
Main Results and Discussion
Automated medium-throughput preparations achieved 87.7 percent recovery relative to manual runs, corresponding to 12.3 percent deviation in total area and 0 percent deviation in relative glycoform distribution. The high-throughput protocol matched medium-throughput manual performance with 23.0 percent deviation in total area and 0 percent in relative area. Intra-run reproducibility remained robust: total area RSD was up to 11.6 percent for 24-sample runs and 15.2 percent for 48-sample runs, while relative area RSD averaged 0.5 percent.
Benefits and Practical Applications
- Increased sample throughput for glycan profiling
- Reduced hands-on time and user variability
- Cost-effective automation with existing laboratory robotics
- Integrated fluorescent and MS detection sensitivity
- Scalable protocols suitable for QA/QC and bioprocess development
Future Trends and Possibilities
Advances in liquid-handling platforms and software connectivity will expand automation of complex glycomics assays. Integration with laboratory information management systems, AI-driven protocol optimization and further scaling beyond 96-well formats are anticipated. Such trends will enhance data consistency, reduce operator variability and accelerate biopharmaceutical development.
Conclusion
The Andrew+-automated GlycoWorks RapiFluor-MS protocols for 24- and 48-sample N-glycan analysis demonstrate comparable performance to manual methods, meeting stringent recovery and reproducibility criteria. These workflows offer high efficiency, robust data quality and straightforward method transferability for both medium- and high-throughput laboratories.
Reference
- Fournier J. A Review of Glycan Analysis Requirements. 2015.
- Dahodwala H, Sharfstein ST. Biosimilars: Imitation Games. ACS Publications; 2017.
- Lauber MA, Yu YQ, Brousmiche DW, et al. Rapid Preparation of Released N-Glycans for HILIC Analysis Using a Labeling Reagent That Facilitates Sensitive Fluorescence and ESI-MS Detection. Anal Chem. 2015;87(10):5401–5409.
- Koza SM, McCall SA, Lauber MA, Chambers EE. Quality Control and Automation Friendly GlycoWorks RapiFluor-MS N-Glycan Sample Preparation. Waters Corporation; May 2020.
- Reed CE, Fournier J, Vamvoukas N, Koza SM. Automated Preparation of MS-Sensitive Fluorescently Labeled N-Glycans with a Commercial Pipetting Robot. SLAS Technol Transl Life Sci Innov. 2018;23(6):550–559.
- Reed CE, Koza SM, Calciano S. GlycoWorks RapiFluor-MS Automation Using the Andrew+ Pipetting Robot. Waters Corporation; August 2020.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.