Sample Preparation, LC/TOF, LC/HRMS, LC/MS, LC/MS/MS, LC/QQQ
IndustriesClinical Research
ManufacturerAgilent Technologies
Význam tématu
Analyses of phosphoproteomes provide insights into cell signaling pathways and disease mechanisms. Reliable enrichment of low-abundance phosphopeptides is essential for reproducible mass spectrometry–based phosphoproteomics in research and quality control.
Cíle a přehled studie
This study evaluated the performance and reproducibility of an automated phosphopeptide enrichment workflow using Agilent AssayMAP Bravo with high-capacity Fe(III)-NTA cartridges. By varying sample/resin ratios (R1–R4) and spiking with light and heavy isotopic phosphopeptide mixes, the aim was to assess enrichment yield, selectivity, and identification depth.
Použitá metodika a instrumentace
- Sample preparation: MCF7 cell line tryptic digests spiked with 20 light phosphopeptide standards.
- Enrichment: Agilent AssayMAP Bravo v2.0 with Fe(III)-NTA cartridges, four distinct sample-to-resin ratios.
- Discovery proteomics: Agilent 1290 Infinity II UHPLC with Nanodapter to nanoflow, coupled to 6550 iFunnel Q-TOF MS; data-dependent acquisition; identification by Spectrum Mill at 1.5% FDR.
- Quantitation: Agilent 6495B triple quadrupole LC/MS with MRM transitions optimized via Skyline; heavy standards spiked post-enrichment for yield calculation.
Hlavní výsledky a diskuse
- Enrichment selectivity improved from ~1.5% before to over 90% after across all ratios.
- Distinct phosphopeptide identifications ranged from ~1,513 (R1) to ~1,188 (R4), increasing with higher sample loads though not strictly proportional.
- Average enrichment yield of the 16 detected standards was ~63% with <5% RSD across ratios, demonstrating high reproducibility.
- Individual peptide yields varied (4–100%), reflecting sequence-dependent binding efficiencies.
Přínosy a praktické využití metody
The automated workflow reduces manual variability, enhances throughput, and delivers consistent high selectivity, supporting both sensitive discovery and targeted quantitative phosphoproteomics. Integration with nanoflow LC/MS and Skyline streamlines method development and data processing.
Budoucí trendy a možnosti využití
- Miniaturization and higher-capacity cartridges for lower sample inputs.
- Automated bioinformatics pipelines for real-time data analysis.
- Multiplexed quantitation with novel isotopic labeling strategies.
- Applications in clinical, single-cell, and spatial phosphoproteomics.
Závěr
The Agilent AssayMAP Bravo automated enrichment workflow with Fe(III)-NTA cartridges offers robust, reproducible phosphopeptide enrichment from complex cell digests across variable sample loads. Coupled with nanoflow LC–MS and automated MRM quantitation in Skyline, it provides a versatile platform for comprehensive phosphoproteomic studies.
Reference
- Wu S, Wu L. Agilent AssayMAP Bravo enables reproducible automated phosphopeptide enrichment using Fe(III)-NTA cartridges. Agilent Technologies Application Note 5991-6073EN (2018).
- Phosphomix peptide standards. Sigma-Aldrich, 2018.
- Agilent Nanodapter for discovery proteomics using nanoflow LC/MS. Application Note 5991-8174EN (2017).
- Agilent 6495 Triple Quadrupole LC/MS: Peptide quantitation performance. Application Note 5991-6898EN (2017).
- Agilent triple quadrupole LC/MS peptide quantitation with Skyline. Application Note 5990-9887EN (2016).
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.