The Agilent 1200 Series High Performance Autosampler G1367C (also referred to as the SL model) is specifically designed to enhance the capabilities of the Agilent 1200 Series Rapid Resolution LC system. This model offers increased pressure range, enabling the use of advanced column technologies such as sub-two-micron columns, which are essential for high-resolution and rapid analyses
Technical Specifications:
- Injection Volume Range: 0.1 – 100 µL in 0.1 µL increments; up to 1500 µL with multiple draw (hardware modification required).
- Injection Precision:
- < 0.25% RSD for 5 – 100 µL.
- < 1% RSD for 1 – 5 µL.
- Pressure Range: 0 – 600 bar (0 – 8700 psi).
- Sample Viscosity Range: 0.2 – 5 cP.
- Sample Capacity:
- 2 × well plates (MTP) + 10 × 2 mL vials.
- 108 × 2 mL vials in 2 × 54 vial plates plus 10 additional 2 mL vials.
- 30 × 6 mL vials in 2 × 15 vial plates plus 10 additional 2 mL vials.
- 54 Eppendorf tubes (0.5/1.5/2.0 mL) in 2 × 27 Eppendorf tube plates.
- Injection Cycle Time: Typically < 17 seconds using standard conditions (default draw speed: 100 µL/min; default eject speed: 100 µL/min; injection volume: 1 µL).
- Carryover: Typically < 0.01% under specified conditions.
- Communications: Controller-area network (CAN), RS-232C, APG-remote standard; optional four external contact closures and BCD vial number output.
- Safety Features: Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display.
- GLP Features: Early maintenance feedback (EMF), electronic records of maintenance and errors.