The Agilent 1290 Infinity HPLC Autosampler Thermostat (G1330B) is an advanced temperature-control module designed to maintain precise thermal conditions for high-performance liquid chromatography (HPLC) applications. The thermostat enhances the accuracy and reproducibility of analytical results by maintaining the optimal temperature for samples, even in variable laboratory conditions. This efficient, Peltier-based system ensures minimal sample degradation and delivers reliable performance across a broad range of applications, making it an essential tool for analytical laboratories.
The G1330B is capable of both heating and cooling, providing a temperature range suitable for most chromatographic needs. It is equipped with multiple safety and diagnostic features to ensure consistent operation and is compatible with various configurations, offering flexibility to researchers and technicians.
Technical Specifications
General Features
- Compatibility: Designed for use with Agilent 1200 Infinity Series or other LC systems with adequate remote control inputs/outputs
- Thermostatted Sample Tray Capacity: Supports either 100 × 1.8 mL vials or two well plates with additional 10 × 1.8 mL vials
Temperature Control
- Temperature Range: Settable from 4 °C to 40 °C in 1 °C increments
- Temperature Accuracy:
- ±3 °C to ±8 °C at ambient temperature <25 °C, humidity <50%
- ±3 °C to ±9 °C at ambient temperature <30 °C, humidity <60%
Physical Characteristics
- Weight: 20.7 kg (46 lbs)
- Dimensions (H × W × D): 140 × 345 × 435 mm (5.5 × 13.5 × 17 inches)
Power Requirements
- Line Voltage: 100 – 240 VAC, ±10%
- Line Frequency: 50 or 60 Hz, ±5%
- Power Consumption: 260 VA / 210 W / 717 BTU
Operating Conditions
- Ambient Operating Temperature: 4 °C to 40 °C (39 °F to 104 °F)
- Humidity: <95% relative humidity, non-condensing
- Altitude: Up to 2000 m (6562 ft) for operation; up to 4600 m (15091 ft) for storage
Safety and Compliance
- Standards: IEC, CSA, UL
- Installation Category: II
- Pollution Degree: 2 (for indoor use only)
Additional Features
- Cooling Mechanism: Peltier-controlled heat exchangers with fans for consistent air circulation
- Condensation Management: Condensation is directed to the leak system for safe disposal
- Maintenance Features: Easily accessible power supply fuses and modular design for simple upkeep
Connectivity
- Control Interfaces: Operable via Agilent Instant Pilot or Agilent control software for LC systems
- Cable Types: Includes connection for autosampler, power, and CAN/LAN communication cables