The Waters 600E Multisolvent Delivery Pump is a key component of the Waters 600E High-Performance Liquid Chromatography (HPLC) system, designed for precise eluent delivery in both isocratic and gradient separations. Its sophisticated design incorporates advanced features like Auto•Blend™ for gradient formation, helium sparging for effective degassing, and a variable flow rate, providing versatility and precision for analytical and diagnostic applications. Ideal for use in complex chromatography workflows, the pump delivers high reliability and reproducibility, making it indispensable in modern laboratories.
Technical Specifications
Flow Rate:
- Adjustable range: 0.01 to 45.0 mL/min
- Increment: 0.01 mL/min
Maximum Pressure:
- 6000 psi (420 kg/cm²) at 10 mL/min
- 1000 psi (70 kg/cm²) at 45 mL/min
Pump Head Volume:
- Standard: 225 µL
- Optional: 100 µL
Gradient Capability:
- Supports up to four eluents via Auto•Blend™
- Preprogrammed gradient profiles: linear, concave, convex
Eluent Degassing:
- Automatic via helium sparging
Flow Path Material:
- Polymeric and stainless steel with low CO₂/O₂ permeability
Interface and Communication:
- IEEE-488 and RS-232 for integration with external devices and data systems
Injector Compatibility:
- Rheodyne 7725i variable-volume manual injector (optional)
Power Requirements:
- Operates on standard AC voltage with safety fuse protection
Operating Environment:
- Temperature: 10–35°C
- Humidity: <80% non