The Agilent 1260 Infinity II HPLC Multisampler G7167A is an advanced, compact, and versatile sampling solution designed to enhance the efficiency and flexibility of high-performance liquid chromatography (HPLC) systems. Capable of handling both vials and microtiter plates, this multisampler ensures seamless operations at pressures of up to 800 bar. With the capacity to accommodate up to 6144 samples, it offers unmatched sample processing capabilities within a single system. Equipped with a dual-needle injection mechanism, the device reduces cycle times and minimizes carryover to as low as 9 ppm through its innovative multiwash feature. Additionally, an integrated sample thermostat is available, enabling precise temperature control from 4°C to 40°C, ensuring sample stability.
The 1260 Infinity II Multisampler is ideal for laboratories requiring high throughput, operational precision, and reduced downtime in chromatographic analyses. Its automated functionality, intuitive control interface, and scalable configuration options make it suitable for a wide range of HPLC applications, from small molecules to biomolecular studies.
Technical Specification
Physical Specifications:
- Weight: 22 kg (without sample thermostat)
- Dimensions: 320 mm (H) × 396 mm (W) × 468 mm (D)
- Line Voltage: 100 – 240 V~, ±10%
- Line Frequency: 50 or 60 Hz
- Power Consumption: 180 VA, 180 W
- Operating Temperature: 4°C – 40°C
- Non-operating Temperature: -40°C – 70°C
- Humidity: <95% relative humidity (non-condensing)
- Operating Altitude: Up to 3000 m
Performance Specifications:
- Injection Range: 0.1 µL – 900 µL (scalable with hardware modifications)
- Injection Precision: <0.15% RSD for single-needle; <0.2% RSD for dual-needle
- Carryover: <0.009% with multiwash
- Pressure Range: Up to 800 bar
- Sample Capacity:
- 1H drawer: Up to 8 drawers, 16 positions
- 2H drawer: Up to 4 drawers, 8 positions
- 3H drawer: Up to 2 drawers, 4 positions
- Cycle Time: <10 seconds per injection
Additional Features:
- Dual-Needle Design: Enables overlapping injections to reduce cycle time.
- Multiwash System: Reduces carryover with up to three different solvents.
- Sample Thermostat: Optional, providing 4°C to 40°C temperature control.
- Instrument Control: Compatible with LC and CE Drivers A.02.10 or higher.
- Communication Ports: CAN and LAN