The Agilent 1260 Infinity Variable Wavelength Detector (VWD) G1314B is a high-performance detector designed for liquid chromatography (HPLC) applications. Renowned for its reliability, the detector offers precise detection of analytes across a wavelength range of 190 to 600 nm. Its double-beam photometer configuration ensures high accuracy and sensitivity, making it an indispensable tool in HPLC laboratories. The system incorporates a deuterium lamp for consistent UV light intensity, providing stable and reproducible results. Designed with user convenience in mind, the G1314B features easy front access for lamp and flow cell replacements, as well as a built-in holmium oxide filter for wavelength accuracy verification. This detector is compatible with Agilent ChemStation software for seamless data analysis and offers extensive diagnostics and safety features, making it suitable for Good Laboratory Practice (GLP) environments.
Technical Specifications
Detection System
- Type: Double-beam photometer
- Light Source: Deuterium lamp
- Wavelength Range: 190 – 600 nm
- Wavelength Accuracy: ± 1 nm (self-calibration with deuterium lines; verification using holmium oxide filter)
- Bandwidth: 6.5 nm (typical)
Performance Metrics
- Noise: < ± 0.5 × 10⁻⁵ AU at 254 nm (ASTM standard)
- Drift: 3 × 10⁻⁴ AU/h at 254 nm
- Linearity: > 2 AU with 5% deviation
Flow Cell Options
- Standard Flow Cell: 10 mm path length, 14 µL volume, max pressure 40 bar (580 psi)
- High-Pressure Flow Cell: 10 mm path length, 14 µL volume, max pressure 400 bar (5800 psi)
- Micro Flow Cell: 5 mm path length, 1 µL volume, max pressure 40 bar (580 psi)
- Semi-Micro Flow Cell: 6 mm path length, 5 µL volume, max pressure 40 bar (580 psi)
Data Handling and Connectivity
- Data Rate: 13 Hz
- Analog Outputs: 100 mV or 1 V (output range 0.001 to 2 AU)
- Communication Interfaces: CAN, RS-232C, optional LAN interface
Environmental Requirements
- Operating Temperature: 0 – 55 °C (32 – 131 °F)
- Humidity: < 95% non-condensing at 40 °C (104 °F)
- Operating Altitude: Up to 2000 m (6562 ft)
Safety and Maintenance
- Built-in leak detection and safe handling of leaks
- Low-voltage areas for maintenance
- Early Maintenance Feedback (EMF) for tracking lamp usage and error logs
Physical Dimensions
- Weight: 11 kg (25 lbs)
- Dimensions: 140 mm (H) × 345 mm (W) × 435 mm (D)