The Agilent 1200 Series Refractive Index Detector (RID) G1362A is a high-performance detector designed for applications requiring precise and reliable refractive index measurements. This specialized detector is ideal for the analysis of non-UV-absorbing compounds, such as carbohydrates, lipids, and polymers, commonly used in industries such as food, pharmaceuticals, and chemicals. By leveraging advanced optical technologies and a robust design, the G1362A ensures accurate results across a broad range of operating conditions.
The detector offers features such as automatic solvent recycling, optical zeroing, and precise temperature control for the optics, ensuring stable baselines and low noise levels. Integrated with the Agilent 1200 Series Control Module and supported by Agilent ChemStation software, it provides a user-friendly interface for method development, data evaluation, and diagnostics. Its rugged build and advanced diagnostics make it a reliable choice for high-throughput laboratories.
Technical Specification
- Detection Type: Refractive index measurement
- Refractive Index Range: 1.00 to 1.75 RIU (calibrated)
- Measurement Range: ±600 x 10⁻⁶ RIU
- Optical Zeroing: Via set screw
- Optics Temperature Control: 5°C above ambient to 55°C
- Sample Cell Volume: 8 µL
- Maximum Pressure: 5 bar (0.5 MPa)
- Maximum Flow Rate: 5 mL/min
- Valves: Automatic purge and solvent recycle
- Liquid Contact Materials: 316 stainless steel, Teflon, quartz glass
- pH Range: 2.3 – 9.5
- Performance:
- Short-term noise: < ±2.5 x 10⁻⁹ RIU
- Drift: < 200 x 10⁻⁹ RIU/hour
- Programmable Parameters: Polarity, peak width
- Control & Data Evaluation:
- Agilent 1200 Series Control Module
- Agilent ChemStation software for PC-based control and evaluation
- Optional PCMCIA card for storage and transfer
- Analog Outputs: 100 mV or 1 V (selectable)
- Communication Interfaces: CAN, GPIB, RS-232C, LAN, APG remote signals
- Dimensions: 345 × 435 × 180 mm (13.5 × 17 × 7 inches)
- Weight: 17 kg (38 lbs)
- Power Requirements:
- Voltage: 100–240 VAC ±10%
- Frequency: 50 or 60 Hz ±5%
- Consumption: 160 VA max
- Operating Conditions:
- Temperature: 0–55°C
- Humidity: < 95%, non-condensing
- Altitude: Up to 2000 m (operating), 4600 m (non-operating)