Superior Solvent Delivery Performance
With an automatic pulsation-correction mechanism and high-speed micro plunger driving, LC-20AD achieves pulse-free solvent delivery. Thanks to improvements in solvent-delivery control firmware, solvent-delivery performance in the micro-flow-rate range below 50 μL/min has been significantly improved.
Low-Pressure Gradient Unit
The optional low-pressure gradient unit can be incorporated in the LC-20AD, enabling gradient elution in a compact space with a small void volume. Automatic matching adjustment of the solenoid valve and pump gives concentration accuracies of ±1.0%
A new type of plunger made with a technique that reflects consideration of the material structure and a precise plunger-holding mechanism help to increase the service life of plunger seals and enable stable solvent delivery over long periods.
- Solvent delivery method: Parallel-type double plunger
- Plunger capacity: 10 μL
- Maximum discharge pressure: 40 MPa
- Flow rate setting range: 0.0001 to 10.0000 mL/min
- Flow rate accuracy: No more than ±1% or ±2 μL/min, whichever is greater (0.01 to 2 mL/min)
- Flow rate precision: No more than 0.06% RSD or 0.02 min SD, whichever is greater
- Gradient type: High-pressure mixing/low-pressure mixing
- Mixing-concentration precision: 0.1% RSD max.
- Constant-pressure solvent delivery: Supported
- Plunger rinsing mechanism: Manual rinsing or automatic rinsing using optional product
- Safety measures: Liquid-leakage sensor, high-pressure/low-pressure limits
- Operating temperature range: 4 to 35°C
- Dimensions, weight: W260 × D420 × H140 mm, 10 kg
- Power requirements: AC 110 V, 230 V, 150 VA, 50/60 Hz