The Agilent 8890 GC System (G3540A) is a gas chromatograph designed for qualitative and quantitative analysis of gaseous mixtures. The offered system is equipped with a valve system for gas sampling, flame ionization detector (FID), thermal conductivity detector (TCD), two Agilent J&W GC columns, control computer, and chromatography software.
The Agilent 8890 gas chromatograph features sixth-generation electronic pneumatics control (EPC) for precise regulation of carrier gas pressure and flow. The instrument includes an integrated touchscreen for analytical method configuration, operating parameter monitoring, and system diagnostics. Its valve configuration supports automated gas sample introduction and control of chromatographic flow paths according to the specific system arrangement.
The combination of FID and TCD detectors enables the analysis of different groups of components in gaseous mixtures. The FID is primarily used for detecting organic compounds, particularly hydrocarbons, while the TCD provides detection of permanent gases and other compounds based on differences in thermal conductivity relative to the carrier gas.
The system includes two capillary columns: Agilent J&W HP-PLOT Q (19095P-QO4PT) and Agilent J&W HP-PLOT Molesieve (19095P-MS0). The HP-PLOT Q column is suitable for separating carbon dioxide and light hydrocarbons, while the HP-PLOT Molesieve column enables separation of permanent gases, including oxygen, nitrogen, and carbon monoxide. This column combination is suitable for the analysis of industrial and process gases, gaseous fuels, and other gas mixtures.
Technical Specifications
- Model: 8890 Gas Chromatograph
- Model Number: G3540A
- Configuration: Valve system, FID, TCD, 2 GC columns, PC, software
- Sample Introduction: Valve-based gas sampling system
- Valve Module: Valve compartment with populated positions 1 and 2
- Flow and Pressure Control: Sixth-generation Electronic Pneumatics Control (EPC)
- EPC Pressure Setting Precision: Resolution down to 0.001 psi
- Carrier Gas Control Modes: Constant pressure, programmed pressure, constant flow, programmed flow
Detector 1: Flame Ionization Detector (FID)
- FID Minimum Detectable Level: <1.2 pg C/s (tridecane, under specified conditions)
- FID Linear Dynamic Range: >10⁷
- Maximum FID Temperature: 450 °C
Detector 2: Thermal Conductivity Detector (TCD)
- TCD Minimum Detectable Level: 400 pg tridecane/mL with helium carrier gas
- TCD Linear Dynamic Range: >10⁵
- Maximum TCD Temperature: 400 °C
GC Oven
- GC Oven Temperature Range: Ambient +4 °C to 450 °C
- Oven Temperature Programming: Up to 20 temperature ramps and 21 plateaus
- Maximum Oven Heating Rate: 120 °C/min depending on configuration and power supply
GC Column 1: Agilent J&W HP-PLOT Q PT
- Part Number: 19095P-QO4PT
- Type: PLOT capillary GC column
- Dimensions: 30 m × 0.53 mm × 40 µm
- Stationary Phase: HP-PLOT Q
- Temperature Range: −60 to 270/290 °C
- Configuration: Two particle traps (front and rear)
GC Column 2: Agilent J&W HP-PLOT Molesieve
- Part Number: 19095P-MS0
- Type: Molecular sieve PLOT capillary GC column
- Dimensions: 30 m × 0.53 mm × 50 µm
- Stationary Phase: HP-PLOT Molesieve
- Temperature Range: −60 to 300 °C
Instrument Control
- Instrument Control: Integrated 7-inch touchscreen and computer control
- Communication Interface: LAN
- Control Computer: PC
- Software: Chromatography software for GC instrument control and data analysis
Applications: Analysis of permanent gases, light hydrocarbons, industrial and process gases, and gaseous fuels