Agilent Cary 3500 Compact UV-VIS Module, G9893A (New)

Agilent Cary 3500 Compact G9893A is a two-position UV-Vis sample/reference measurement module. Combined with the G9864A Engine, it provides 190–1100 nm measurements, scanning up to 150,000 nm/min, and data collection at 250 points/s.
Purchase price
3,500
€
without VAT
In stock
+ 100 € delivery

Features

Condition:
New
Warranty:
3 months
Manufacture year:
2023
Countries:
Czech Republic
Additional services
Installation
Basic training
Extended warranty

Features

Agilent Cary 3500 Compact UV-Vis Module, G9893A, is a compact two-position measurement module for the modular Agilent Cary 3500 UV-Vis spectrophotometer platform. It is designed for simultaneous cuvette-based sample and reference measurement and, when combined with the Cary 3500 UV-Vis Engine, provides fast and accurate UV-Vis measurements over the 190–1100 nm wavelength range.

G9893A is the measurement module only. It does not contain the light source or monochromator; these components are located in the separate Cary 3500 UV-Vis Engine G9864A. The module is designed for ambient measurements and, unlike G9894A, does not incorporate Peltier temperature control.

Cary 3500 Modular Architecture

The Cary 3500 uses a modular architecture consisting of two primary components:

  • Cary 3500 UV-Vis Engine G9864A
  • a measurement module such as the G9893A Compact UV-Vis Module

The Engine contains:

  • the full-spectrum xenon flash lamp,
  • monochromator,
  • primary optical components.

G9893A provides the sample and reference measurement compartment and connects directly to the UV-Vis Engine.

This architecture allows different measurement modules to be used with the same optical engine.

Two Cuvette Positions

The Compact UV-Vis Module is designed for simultaneous measurement of:

  • 1 sample
  • 1 reference

The two-channel configuration supports double-beam UV-Vis measurement and continuous correction for changes in light-source intensity.

The module supports conventional cuvette measurements as well as low-volume measurements when appropriate cuvettes are used.

Optical Design

Cary 3500 measurement modules contain no moving optical parts and feature permanent optical alignment.

The optical system provides a tightly controlled beam measuring less than approximately 1.5 mm at the sample interface.

This supports measurements using:

  • conventional cuvettes,
  • microvolume cuvettes,
  • small-aperture cells,
  • limited sample volumes.

UV-Vis Performance of the Complete System

When coupled to the Cary 3500 UV-Vis Engine, the system provides:

  • Wavelength Range: 190–1100 nm
  • Spectral Bandwidth: 0.1–5.0 nm
  • Bandwidth Increment: 0.01 nm
  • Maximum Scan Rate: 150,000 nm/min
  • Data Collection Rate: up to 250 points/s
  • Photometric Range: up to 4 Abs
  • Limiting Resolution: 0.1 nm
  • Wavelength Accuracy: ±0.2 nm
  • Wavelength Reproducibility: <0.025 nm
  • Photometric Accuracy: ±0.005 Abs at 1 Abs

These specifications apply to the complete Cary 3500 UV-Vis system with the appropriate Engine.

Light Source

The light source is not contained in G9893A.

The Cary 3500 UV-Vis Engine uses a full-spectrum xenon flash lamp operating at 250 Hz.

The xenon source supports:

  • UV and visible measurements using a single source,
  • fast data acquisition,
  • rapid kinetic measurements,
  • reduced unnecessary sample exposure between measurements.

Technical Specifications

  • PN: G9893A
  • Technique: UV-Vis Spectrophotometry
  • Product Type: Measurement Module
  • Measurement Positions: 2
  • Configuration: Sample + Reference
  • Temperature Control: No
  • Operating Configuration: Ambient measurements
  • Compatible Optical Engine: Cary 3500 UV-Vis Engine G9864A
  • Complete System Wavelength Range: 190–1100 nm
  • Spectral Bandwidth: 0.1–5.0 nm
  • Maximum Scan Rate: 150,000 nm/min
  • Data Collection Rate: up to 250 points/s
  • Photometric Range: 4 Abs
  • Complete Instrument Optical System: Double beam
  • Beam Size at Sample Interface: < 1.5 mm
  • Module Height: 280 mm
  • Module Width: 420 mm
  • Module Depth: 205 mm
  • Module Weight: 6.6 kg
  • Software: Cary UV Workstation depending on configuration