Thermo Dionex DGP-3600 Dual Low-Pressure Gradient HPLC Pump, DGP-3600AB (Refurbished)

Thermo Scientific Dionex DGP-3600AB is a biocompatible UltiMate 3000 dual ternary HPLC pump with two independent gradients, flow rates up to 10 mL/min, and pressure up to 500 bar.
Purchase price
1,850
€
without VAT
In stock
+ 150 € delivery

Features

Condition:
Refurbished
Warranty:
3 months
Manufacture year:
0
Countries:
Czech Republic
Additional services
Installation
Extended warranty
Basic training

Features

Thermo Scientific Dionex UltiMate 3000 DGP-3600AB is a biocompatible dual low-pressure gradient HPLC pump designed for advanced analytical LC applications. It combines two independent ternary gradient pumps in a single module, enabling dual-flow-path operation, tandem workflows, 2D-LC, and automated sample-preparation applications.

The DGP-3600AB features a biocompatible flow path with titanium components. Each pump uses a serial dual-piston design and provides low-pressure proportioning of three solvents, giving a total of six eluent lines (2 × 3).

Two Gradient Pumps in One Module

The DGP-3600AB contains two separate pumping units that can operate independently. Each unit provides three solvent channels and generates ternary gradients through low-pressure proportioning.

This architecture supports applications including:

  • parallel LC separations,
  • tandem chromatographic workflows,
  • 2D-LC,
  • online SPE and enrichment,
  • automated matrix elimination,
  • use of one pump for separation and the second for sample preparation or transfer.

Biocompatible Design

The AB designation identifies the biocompatible analytical version.

Fluidic components use materials designed to minimize contact of the sample with stainless steel. Wetted materials include components manufactured from materials such as:

  • titanium,
  • PEEK,
  • PTFE,
  • FEP,
  • ECTFE,
  • sapphire,
  • zirconium oxide,
  • aluminum oxide.

This configuration is suitable for biochromatography and applications involving sensitive biomolecules.

Flow and Pressure

The analytical DGP-3600AB is specified for:

  • recommended flow range: 0.2–10.0 mL/min
  • settable flow range: 0.001–10.0 mL/min
  • pressure range: 2–50 MPa
  • maximum pressure: 500 bar / 7250 psi

Flow accuracy is ±0.1% at 1 mL/min.

Flow precision is specified as:

  • <0.05% RSD or <0.01 min SD, whichever is greater.

Gradient Proportioning

Each pumping unit generates a gradient by low-pressure solvent proportioning.

  • Number of eluent lines: 6 (2 × 3)
  • Proportioning Accuracy: ±0.5% of full scale
  • Proportioning Precision: <0.2% SD at 2 mL/min
  • Gradient Delay Volume: approximately 690 µL

The standard dynamic mixing chamber has a volume of approximately 328 µL and can be expanded using optional mixing-chamber extension kits.

Flow and Pressure Stability

The serial dual-piston design and electronic flow control help minimize pressure pulsation.

Typical pressure ripple is:

  • <0.2 MPa or <1%, whichever is greater.

Stable flow and gradient formation support reproducible retention times and quantitative LC performance.

Solvent Degassing

The DGP-3600AB does not have an integrated solvent degasser.

An external UltiMate 3000 Solvent Rack with an appropriate degassing system can be used when solvent degassing is required.

This is especially useful for gradient applications where dissolved gases can affect flow and gradient stability.

Active Rear-Seal Wash

The pump incorporates an active rear-seal wash system.

This supports:

  • extended piston-seal lifetime,
  • removal of mobile-phase deposits,
  • operation with buffered mobile phases,
  • monitoring of piston-seal integrity.

Communication and Control

The DGP-3600AB provides:

  • USB PC connection,
  • integrated three-port USB hub,
  • 3 digital inputs,
  • 4 programmable relay outputs,
  • analog pressure-monitoring output,
  • 15-pin D-Sub connection for the Solvent Rack / degasser.

The front panel includes an LCD, status LEDs, and function keys for installation and maintenance.

Chromeleon CDS

The pump can be integrated with Thermo Scientific Chromeleon™ Chromatography Data System (CDS).

Supported functions can include:

  • control of both pumps,
  • gradient programming,
  • pressure monitoring,
  • system diagnostics,
  • System Wellness monitoring,
  • Automatic Equipment Qualification (AutoQ),
  • logging of system parameters in the Audit Trail.

Specific capabilities depend on the installed CDS version and license.

Technical Specifications

  • Operating Principle: Serial Dual-Piston
  • Number of Pump Units: 2
  • Gradient Formation: Dual Low-Pressure Proportioning
  • Number of Eluent Lines: 6 (2 × 3)
  • Recommended Flow Range: 0.2–10.0 mL/min
  • Settable Flow Range: 0.001–10.0 mL/min
  • Flow Accuracy: ±0.1% at 1 mL/min
  • Flow Precision: <0.05% RSD or <0.01 min SD
  • Pressure Range: 2–50 MPa
  • Maximum Pressure: 500 bar / 7250 psi
  • Pressure Ripple: typically <0.2 MPa or <1%
  • Proportioning Accuracy: ±0.5% of full scale
  • Proportioning Precision: <0.2% SD at 2 mL/min
  • Gradient Delay Volume: 690 µL
  • Solvent Degassing: External, optional
  • Wetted Flow Path: Biocompatible with titanium components
  • Rear Seal Wash: Active
  • PC Connection: USB
  • Integrated USB Hub: 3 ports
  • Dimensions (H × W × D): 19 × 42 × 51 cm
  • Weight: 21.5 kg
  • Power: 100–120 V / 60 Hz or 200–240 V / 50 Hz
  • Power Consumption: max. 150 VA
  • Software: Thermo Scientific Chromeleon™ CDS depending on version and license