Agilent 6150B Single Quadrupole LC/MS (G6150B) is a benchtop single-quadrupole mass spectrometer designed for rapid LC/MS identification, compound confirmation, and quantitative analysis. It combines fast scanning, rapid polarity switching, and compatibility with multiple atmospheric-pressure ionization sources, making it suitable for routine analysis, method development, and fast UHPLC separations.
The 6150B provides a mass range up to m/z 1350, scan speeds up to 10,000 u/s, and positive-to-negative polarity switching in approximately 20 ms. It supports Full Scan and SIM acquisition with unit mass resolution.
Single Quadrupole Mass Analyzer
The 6150B uses a quadrupole mass filter to select ions according to their mass-to-charge ratio.
Typical acquisition modes include:
- Full Scan – acquisition of a mass spectrum across a defined mass range
- SIM – Selected Ion Monitoring – targeted monitoring of selected ions for increased sensitivity
Single-quadrupole LC/MS is particularly useful for rapid molecular-weight confirmation, targeted compound monitoring, and routine quantitative analysis without the need for tandem MS/MS.
Fast Acquisition
A maximum scan speed of up to 10,000 u/s provides sufficient data density for narrow chromatographic peaks generated by rapid LC and UHPLC separations.
This is beneficial for:
- fast gradient methods,
- high-throughput screening,
- UHPLC separations,
- short-retention-time methods,
- rapid qualitative LC/MS workflows.
Rapid Polarity Switching
The 6150B can switch between positive and negative ionization in approximately 20 ms.
This enables compounds responding in different ionization polarities to be monitored during the same analysis without requiring separate injections.
Ionization Sources
The Agilent 6150B supports interchangeable ionization sources, depending on system configuration, including:
- ESI – Electrospray Ionization
- Agilent Jet Stream ESI
- APCI – Atmospheric Pressure Chemical Ionization
- APPI – Atmospheric Pressure Photoionization
- Multimode ESI/APCI
Agilent Jet Stream is supported on the 6150B and can provide enhanced ESI sensitivity. The ion source is configuration-dependent and should not be assumed to be included with the G6150B mainframe itself.
Sensitivity
Agilent specifies high positive-ESI SIM sensitivity for the 6150B using reserpine under defined test conditions.
Actual system sensitivity depends on factors including:
- installed ion source,
- condition of the ion optics and detector,
- mobile-phase flow rate,
- sample matrix,
- source and fragmentor settings.
Software
The G6150B can be controlled using Agilent LC/MS ChemStation / OpenLab ChemStation, depending on the installed software version and license.
Current Agilent documentation continues to list the G6150B as a supported Single Quadrupole LC/MS module in OpenLab ChemStation LTS.
Software capabilities include:
- LC/MS system control,
- acquisition method setup,
- Full Scan and SIM acquisition,
- time programming,
- chromatogram and mass-spectrum processing,
- quantitative analysis.
Technical Specifications
- Mass Range: up to m/z 1350
- Maximum Scan Speed: up to 10,000 u/s
- Acquisition Modes: Full Scan, SIM
- Polarity: Positive / Negative
- Polarity Switching: approx. 20 ms
- Mass Resolution: Unit Resolution
- Mass Accuracy: approx. ±0.13 Da
- Mass Axis Stability: ±0.1 u or 100 ppm / 12 h, whichever is larger
- Minimum SIM Dwell Time: 5 ms
- Maximum SIM Ions: up to 100 per time segment
- Dynamic Range: > 6 × 10⁶
- Ion Sources: ESI, Agilent Jet Stream, APCI, APPI, Multimode depending on configuration
- Vacuum System: Turbomolecular + mechanical pump
- Autotune: Yes