The Waters ACQUITY UPLC I-Class System is a high-performance ultra-performance liquid chromatography (UPLC) system designed for high-resolution separations, precise solvent delivery, and reliable sample analysis. The system combines a Binary Solvent Manager (BSM), a Sample Manager with Flow-Through Needle (SM-FTN), and a Tunable UV (TUV) Detector.
The Binary Solvent Manager provides high-pressure binary gradient mixing with integrated vacuum degassing. The Sample Manager-FTN enables automated sample injection with programmable needle washing and temperature-controlled sample storage. The TUV Detector provides UV absorbance detection over a wavelength range of 190–700 nm.
Key Features and Benefits
- High-pressure UPLC performance: Operating pressures up to 18,000 psi (1,241 bar) at flow rates up to 1.0 mL/min.
- Binary gradient delivery: High-pressure mixing of two solvents, with selection from up to four solvent reservoirs.
- Integrated vacuum degassing: Six-channel degasser for mobile phases and sample manager wash solvents.
- Automated sample injection: Flow-through needle technology with programmable needle washing.
- Temperature-controlled sample management: Sample compartment temperature range of 4–40 °C.
- UV absorbance detection: Tunable UV detector with a wavelength range of 190–700 nm.
- System monitoring: Integrated leak detection and diagnostic functions.
Technical Specifications
UPLC System
- Manufacturer: Waters
- Model: ACQUITY UPLC I-Class
- Instrument type: Ultra-performance liquid chromatography (UPLC) system
- Configuration: Binary Solvent Manager, Sample Manager-FTN, TUV Detector
- Operating flow rate range: 0.010–2.000 mL/min
- Maximum operating pressure: 1,241 bar (18,000 psi) up to 1.0 mL/min; 827 bar (12,000 psi) up to 2.0 mL/min
Binary Solvent Manager (BSM)
- Pump type: High-pressure binary gradient pump
- Solvent selection: A1/A2 and B1/B2
- Simultaneously mixed solvents: 2
- Gradient formation: High-pressure mixing
- Degassing: Integrated six-channel vacuum degasser
- Gradient profiles: 11
- Flow accuracy: ±1.0% at 0.5 mL/min
- Compressibility compensation: Automatic
- Plunger seal wash: Programmable, active
Sample Manager-FTN
- Injection system: Flow-Through Needle (FTN)
- Injection volume range: 0.1–10.0 µL (standard configuration)
- Sample capacity: Up to 2 sample plates (96- or 384-well microtiter plates or vial trays).
- Sample compartment temperature: 4–40 °C
- Needle wash: Integrated, active, programmable
TUV Detector
- Detector type: Tunable UV absorbance detector
- Wavelength range: 190–700 nm
Included Components
- Waters ACQUITY UPLC I-Class Binary Solvent Manager (BSM)
- Waters ACQUITY UPLC I-Class Sample Manager-FTN
- Waters ACQUITY UPLC Tunable UV (TUV) Detector