Agilent G2571-67487 is a service electronic module for the second mass-selective quadrupole, MS2/Q3, in Agilent Triple Quadrupole LC/MS systems. Operating at an RF frequency of approximately 1.04 MHz, the assembly controls the quadrupole electrical field used to selectively transmit product ions during LC/MS/MS analysis.
Function in a Triple Quadrupole
A triple-quadrupole mass spectrometer uses three principal analyzer sections:
- Q1 / MS1 – precursor-ion selection
- Q2 – collision cell
- Q3 / MS2 – product-ion selection
G2571-67487 / G2571-60487 controls MS2/Q3.
The Quad Driver generates the RF and control voltages required for operation of the second mass filter, allowing Q3 to transmit the selected product ion to the detector.
MS2 / Q3
During a typical MRM experiment:
- Q1 selects the precursor ion.
- Q2 performs collision-induced fragmentation.
- Q3 / MS2 selects the target product ion.
- The detector records the corresponding MRM transition.
Stable MS2 control is therefore essential for product-ion selectivity and reproducible quantitative performance.
1.04 MHz Quad Driver
The Quad Driver operates with an RF field of approximately 1.04 MHz.
Its functions include:
- RF signal generation,
- quadrupole-field control,
- m/z-selective ion transmission,
- rapid setting changes during MRM acquisition.
Quad Driver vs. Quadrupole
This component is not the physical quadrupole mass analyzer itself. It is the electronic assembly used to control the physical quadrupole mass filter.
The:
- Q3 mass filter,
- Quad Driver electronics
are separate components.