Agilent G2571-67487 is an original service Quad Driver assembly designed for electronic control of the second mass-selective quadrupole, MS2/Q3, in compatible Agilent Triple Quadrupole LC/MS systems. Operating at approximately 1.04 MHz RF frequency, the module generates the high-frequency electrical field required for product-ion selection in tandem mass spectrometry.
Function in a Triple Quadrupole
A triple-quadrupole mass spectrometer uses three principal analyzer sections:
- Q1 / MS1 – precursor-ion selection
- Q2 – collision cell
- Q3 / MS2 – product-ion selection
G2571-67487 / G2571-60487 controls MS2/Q3.
The Quad Driver generates the RF and control voltages required for operation of the second mass filter, allowing Q3 to transmit the selected product ion to the detector.
MS2 / Q3
During a typical MRM experiment:
- Q1 selects the precursor ion.
- Q2 performs collision-induced fragmentation.
- Q3 / MS2 selects the target product ion.
- The detector records the corresponding MRM transition.
Stable MS2 control is therefore essential for product-ion selectivity and reproducible quantitative performance.
1.04 MHz Quad Driver
The Quad Driver operates with an RF field of approximately 1.04 MHz.
Its functions include:
- RF signal generation,
- quadrupole-field control,
- m/z-selective ion transmission,
- rapid setting changes during MRM acquisition.
Quad Driver vs. Quadrupole
This component is not the physical quadrupole mass analyzer itself. It is the electronic assembly used to control the physical quadrupole mass filter.
The:
- Q3 mass filter,
- Quad Driver electronics
are separate components.