Agilent 1220 Infinity Compact HPLC System (Isocratic Pump, Manual Injector, VWD), G4286B (Refurbished)

Agilent 1220 Infinity LC System G4286B is a compact HPLC system with an isocratic pump, manual injection valve, and VWD detector. It supports flow rates of 0.2–10 mL/min and pressures up to 600 bar.
Rental per month
400
€
without VAT
Purchase price
8,500
€
without VAT
In stock
+ 200 € delivery

Features

Condition:
Refurbished
Warranty:
3 months
Manufacture year:
0
Countries:
Czech Republic
Additional services
Advanced training
Installation
Extended warranty
Basic training
Application implementation

Features

The Agilent Technologies 1220 Infinity LC System (G4286B) is a compact HPLC system designed for routine isocratic liquid chromatography applications. This configuration combines an isocratic pump, a manual injection valve, and a variable wavelength detector (VWD) for UV absorbance detection.

The system features a dual-piston, in-series pump with servo-controlled variable stroke technology, providing precise and stable mobile phase delivery. It supports flow rates from 0.2 to 10.0 mL/min and operating pressures up to 600 bar. The manual six-port injection valve is equipped with a 20 µL sample loop, while the VWD offers a wavelength range of 190–600 nm. Built-in diagnostics, leak detection, and Early Maintenance Feedback (EMF) support instrument monitoring and maintenance.

Technical Specifications

General Specifications

  • Model: Agilent 1220 Infinity LC System (G4286B)
  • Configuration: Isocratic pump, manual injector, variable wavelength detector (VWD)
  • Separation Mode: Isocratic HPLC
  • Injection System: Manual six-port Rheodyne injection valve
  • Sample Loop Volume: 20 µL
  • Injection Seal: PEEK

Isocratic Pump

  • Pump Technology: Dual-piston, in-series pump with servo-controlled variable stroke drive
  • Settable Flow Range: 0.001–10.0 mL/min
  • Operating Flow Range: 0.2–10.0 mL/min
  • Flow Precision: ≤0.07% RSD or < 0.02 min SD (whichever is greater, based on retention time at constant room temperature)
  • Flow Accuracy: ±1% or 10 µL/min (whichever is greater)
  • Operating Pressure:
    • Up to 60 MPa (600 bar) at flow rates up to 5 mL/min
    • Up to 20 MPa (200 bar) at flow rates up to 10 mL/min
  • Pressure Pulsation: < 2% amplitude (typically < 1.3%)
  • Compressibility Compensation: User-selectable, based on mobile phase compressibility
  • Recommended pH Range: 1.0–12.5

Variable Wavelength Detector (VWD)

  • Detection Technology: Double-beam UV absorbance photometer
  • Light Source: Deuterium lamp
  • Wavelength Range: 190–600 nm
  • Wavelength Accuracy: ±1 nm
  • Optical Bandwidth: Typically 6.5 nm
  • Maximum Data Rate: 80 Hz
  • Short-Term Noise (ASTM): < ±0.25 × 10⁻⁵ AU at 230 nm
  • Drift: < 1 × 10⁻⁴ AU/h at 230 nm
  • Standard Flow Cell: 14 µL volume, 10 mm optical path length

Physical and Electrical Characteristics

  • Weight: 30 kg
  • Dimensions (H × W × D): 640 × 370 × 420 mm
  • Voltage: 100–240 VAC, ±10%
  • Frequency: 50/60 Hz, ±5%
  • Maximum Power Consumption: 240 VA / 210 W

Connectivity & Control

  • Communication Interfaces:
    • CAN
    • RS-232C
    • APG Remote (ready, start, stop, shutdown signals)
    • LAN
  • Chromatography Software: Agilent ChemStation, Agilent EZChrom Compact
  • Instrument Utilities: Agilent Instrument Utilities, Agilent Lab Advisor

Maintenance & Safety Features

  • Diagnostics: Error detection and monitoring through Agilent diagnostic software
  • Leak Detection: Leak detection, safe leak handling, and shutdown signaling
  • Maintenance Monitoring: Early Maintenance Feedback (EMF) for tracking pump usage and detector lamp operating time
  • Safety Standards: IEC, CSA, UL; Installation Category II, Pollution Degree 2 (indoor use only)

Environmental Conditions

  • Operating Temperature: 4–55 °C
  • Non-operating Temperature: −40 to 70 °C
  • Humidity: < 95% RH at 40 °C (non-condensing)
  • Operating Altitude: Up to 2,000 m
  • Non-operating Altitude: Up to 4,600 m

Note: The G4286B configuration includes manual sample injection, an isocratic pump, and a VWD detector. The standard manual injection valve uses a 20 µL sample loop.