The Agilent Technologies 1260 Infinity Bio-Inert Multisampler (G5668A) is an automated HPLC sampling module designed for precise and reliable injection of samples containing biomolecules such as proteins, monoclonal antibodies, and oligonucleotides. Its metal-free sample flow path, featuring a ceramic needle, PEEK needle seat, and stainless steel-clad PEEK capillaries, prevents direct contact between samples and metal surfaces while supporting operating pressures up to 600 bar.
The multisampler accommodates both vials and microtiter plates, with a maximum capacity of 6,144 samples depending on the drawer configuration. It provides a standard injection range of 0.1–100 µL with high injection precision. Optional Multiwash technology can reduce carryover to less than 9 ppm, while an optional sample thermostat enables temperature control for temperature-sensitive samples.
Technical Specifications
General Specifications
- Model: Agilent 1260 Infinity Bio-Inert Multisampler (G5668A)
- Module Type: Bio-inert HPLC autosampler
- Sample Flow Path: Metal-free
- Injection Needle: Ceramic
- Needle Seat: PEEK
- Capillaries: Stainless steel-clad PEEK
- Standard Injection Range: 0.1–100 µL
- Injection Volume Increments: 0.1 µL
- Injection Precision: < 0.15% RSD or SD < 10 nL (whichever is greater)
- Maximum Operating Pressure: 600 bar
- Sample Viscosity Range: 0.2–5.0 cP
- Maximum Sample Capacity: 6,144 samples (16 × 384-well microtiter plates)
- Sample Containers: Vials, microtiter plates, and compatible sample tubes
- Carryover: < 30 ppm (standard); < 9 ppm with optional Multiwash
- Multiwash: Optional needle wash and needle seat backflush using up to three solvents
- Sample Temperature Control: Optional sample thermostat
Physical and Electrical Characteristics
- Weight: 22 kg (without optional accessories)
- Dimensions (H × W × D): 320 × 396 × 468 mm
- Voltage: 100–240 VAC, ±10%
- Frequency: 50/60 Hz, ±5%
- Power Consumption: 180 VA / 180 W
Connectivity & Control
- Communication Interfaces:
- CAN
- LAN
- ERI (ready, start, stop, and shutdown signals)
Maintenance & Safety Features
- Diagnostics: Error detection and monitoring through Agilent Lab Advisor
- Leak Detection: Leak detection, safe leak handling, and shutdown signaling
- Maintenance Monitoring: Early Maintenance Feedback (EMF) for tracking instrument usage with user-configurable limits and maintenance notifications
- Safety Standards: IEC, EN, CSA, UL; Overvoltage Category II, Pollution Degree 2 (indoor use only)
Environmental Conditions
- Operating Temperature: 4–40 °C
- Non-operating Temperature: −40 to 70 °C
- Humidity: < 95% RH at 40 °C (non-condensing)
- Operating Altitude: Up to 3,000 m
Note: Maximum sample capacity, Multiwash functionality, and sample temperature control depend on the installed configuration and optional accessories.