Agilent 1260 Infinity Preparative Pump, G1361A (Refurbished)

Agilent Technologies 1260 Infinity Prep Pump G1361A is a dual-piston pump for preparative HPLC with flow rates up to 100 mL/min. It supports isocratic operation or binary gradients with a second pump.
Purchase price
7,000
€
without VAT
In stock
+ 100 € delivery

Features

Condition:
Refurbished
Warranty:
3 months
Manufacture year:
0
Countries:
Czech Republic
Additional services
Advanced training
Installation
Extended warranty
Basic training
Application implementation

Features

The Agilent Technologies 1260 Infinity Preparative Pump (G1361A) is designed for preparative HPLC applications, providing precisely controlled mobile phase flow rates up to 100 mL/min. Its dual-piston pumping system features two parallel pistons with fixed stroke lengths and a cam-driven mechanism. Overlapping piston delivery cycles and motor speed regulation ensure smooth and stable mobile phase delivery.

The G1361A operates within a pressure range of 20–400 bar and offers user-selectable compressibility compensation for different mobile phases. The pump incorporates an automated seal wash system and an electromagnetic purge valve. It can be used independently for isocratic separations or combined with a second preparative pump to form a high-pressure binary gradient system.

Technical Specifications

General Specifications

  • Model: Agilent 1260 Infinity Preparative Pump (G1361A)
  • Hydraulic System: Dual-piston, parallel configuration with cam-driven, fixed-stroke pistons
  • Settable Flow Range: 0.001–100 mL/min
  • Flow Precision: < 0.5% RSD
  • Operating Pressure: 20–400 bar
  • Compressibility Compensation: User-selectable, based on mobile phase compressibility
  • Recommended pH Range: 1.0–12.5
  • Seal Wash: Integrated automated seal wash system with peristaltic pump
  • Pump Configuration: Isocratic operation or binary gradient operation with an additional preparative pump

Physical and Electrical Characteristics

  • Weight: 15 kg (34 lbs)
  • Dimensions (H × W × D): 440 × 200 × 345 mm
  • Voltage: 100–240 VAC, ±10%
  • Frequency: 50/60 Hz, ±5%
  • Maximum Power Consumption: 250 VA / 210 W

Connectivity & Control

  • Communication Interfaces:
    • CAN
    • RS-232
    • APG Remote (ready, start, stop, shutdown signals)
    • CAN-DC OUT
    • Optional LAN
  • Software Control: Agilent ChemStation for LC and LC/MS

Maintenance & Safety Features

  • Diagnostics: Error detection and monitoring through Agilent ChemStation
  • Leak Detection: Leak detection, safe leak handling, and shutdown signaling
  • Maintenance Monitoring: Early Maintenance Feedback (EMF) for tracking instrument usage, seal wear, and pumped mobile phase volume
  • Safety Standards: IEC, CSA, UL; Installation Category II, Pollution Degree 2 (indoor use only)

Environmental Conditions

  • Operating Temperature: 4–40 °C
  • Non-operating Temperature: −40 to 70 °C
  • Humidity: < 95% RH at 25–40 °C (non-condensing)
  • Operating Altitude: Up to 2,000 m
  • Non-operating Altitude: Up to 4,600 m

Note: For optimal performance, flow rates above 5 mL/min and system pressures above 20 bar are recommended. Solvents with a pH below 2.3 should not contain acids that attack stainless steel.