The Agilent Technologies 1260 Infinity Preparative Pump (G1361A) is a high-performance HPLC pump designed for preparative liquid chromatography applications requiring high solvent flow rates and precise flow control. Its dual-piston, parallel design features cam-driven pistons with fixed stroke lengths and overlapping delivery cycles to ensure smooth and consistent mobile phase delivery.
The pump supports flow rates up to 100 mL/min and operating pressures up to 400 bar. It features user-adjustable compressibility compensation, an integrated automated seal wash system, and an electromagnetic purge valve. The G1361A can operate as a standalone isocratic pump or be combined with a second preparative pump to create a high-pressure binary gradient system.
Technical Specifications
General Specifications
- Model: Agilent 1260 Infinity Preparative Pump (G1361A)
- Hydraulic System: Dual-piston, parallel configuration with cam-driven, fixed-stroke pistons
- Settable Flow Range: 0.001–100 mL/min
- Flow Precision: < 0.5% RSD
- Operating Pressure: 20–400 bar
- Compressibility Compensation: User-selectable, based on mobile phase compressibility
- Recommended pH Range: 1.0–12.5
- Seal Wash: Integrated automated seal wash system with peristaltic pump
- Pump Configuration: Isocratic operation or binary gradient operation with an additional preparative pump
Physical and Electrical Characteristics
- Weight: 15 kg (34 lbs)
- Dimensions (H × W × D): 440 × 200 × 345 mm
- Voltage: 100–240 VAC, ±10%
- Frequency: 50/60 Hz, ±5%
- Maximum Power Consumption: 250 VA / 210 W
Connectivity & Control
- Communication Interfaces:
- CAN
- RS-232
- APG Remote (ready, start, stop, shutdown signals)
- CAN-DC OUT
- Optional LAN
- Software Control: Agilent ChemStation for LC and LC/MS
Maintenance & Safety Features
- Diagnostics: Error detection and monitoring through Agilent ChemStation
- Leak Detection: Leak detection, safe leak handling, and shutdown signaling
- Maintenance Monitoring: Early Maintenance Feedback (EMF) for tracking instrument usage, seal wear, and pumped mobile phase volume
- Safety Standards: IEC, CSA, UL; Installation Category II, Pollution Degree 2 (indoor use only)
Environmental Conditions
- Operating Temperature: 4–40 °C
- Non-operating Temperature: −40 to 70 °C
- Humidity: < 95% RH at 25–40 °C (non-condensing)
- Operating Altitude: Up to 2,000 m
- Non-operating Altitude: Up to 4,600 m
Note: For optimal performance, flow rates above 5 mL/min and system pressures above 20 bar are recommended. Solvents with a pH below 2.3 should not contain acids that attack stainless steel.