Agilent 1260 Infinity HPLC Isocratic Pump (G1310B) (Refurbished)

Agilent Technologies 1260 Infinity Isocratic Pump G1310B features a dual-piston design with low pressure pulsation, flow rates of 0.2–10 mL/min, leak detection, and advanced diagnostics.
Rental per month
200
€
without VAT
Purchase price
5,500
€
without VAT
In stock
+ 100 € delivery

Features

Condition:
Refurbished
Warranty:
3 months
Manufacture year:
0
Countries:
Czech Republic
Additional services
Advanced training
Installation
Extended warranty
Basic training
Application implementation

Features

The Agilent Technologies 1260 Infinity Isocratic Pump (G1310B) is a high-performance isocratic HPLC pump providing precise control of mobile phase flow. Its design incorporates two pistons arranged in series, with a servo-controlled variable stroke mechanism. This configuration delivers stable solvent flow with high precision and reduced pressure pulsation.

The G1310B operates over a flow range of 0.2–10.0 mL/min and supports operating pressures up to 600 bar. It offers user-adjustable compressibility compensation, diagnostic functions, and leak detection. The Early Maintenance Feedback (EMF) system supports instrument usage monitoring and maintenance planning. The pump is designed for isocratic HPLC analysis, where the mobile phase composition remains constant throughout the separation.

Technical Specifications

General Specifications

  • Model: Agilent 1260 Infinity Isocratic Pump (G1310B)
  • Technology: Dual-piston, in-series pump with servo-controlled variable stroke drive
  • Flow Range: 0.2–10.0 mL/min
  • Flow Precision: ≤0.07% RSD or ≤0.02 min SD (whichever is greater)
  • Flow Accuracy: ±1% or 10 µL/min (whichever is greater)
  • Operating Pressure:
    • Up to 60 MPa (600 bar) at flow rates up to 5 mL/min
    • Up to 20 MPa (200 bar) at flow rates up to 10 mL/min
  • Pressure Pulsation: < 2% amplitude (typically < 1.3%) or < 0.3 MPa (3 bar), whichever is greater, measured at 1 mL/min with isopropanol at pressures above 10 bar
  • Compressibility Compensation: User-selectable, based on mobile phase compressibility

Physical and Electrical Characteristics

  • Weight: 11 kg (25 lbs)
  • Dimensions (H × W × D): 180 × 345 × 435 mm
  • Voltage: 100–240 VAC, ±10%
  • Frequency: 50/60 Hz, ±5%
  • Maximum Power Consumption: 180 VA, 55 W / 188 BTU

Connectivity & Control

  • Communication Interfaces: 
    • CAN
    • RS-232C
    • APG Remote (ready, start, stop, shutdown signals)
    • optional LAN
  • Software Control: 
    • Agilent ChemStation
    • EZChrom
    • MassHunter
    • OpenLab
  • Local Control: Agilent Instant Pilot
  • Analog Output: 1.33 mV/bar for pressure monitoring

Maintenance & Safety Features

  • Diagnostics: Error detection and monitoring through Agilent LabAdvisor
  • Leak Detection: Leak detection, safe leak handling, and shutdown signaling
  • Maintenance Monitoring: Early Maintenance Feedback (EMF) for tracking instrument usage, seal wear, and pumped mobile phase volume
  • Safety Standards: IEC, CSA, UL; Installation Category II, Pollution Degree 2 (indoor use only)

Environmental Conditions

  • Operating Temperature: 4–55 °C
  • Non-operating Temperature: -40 to 70 °C
  • Humidity: < 95% RH at 40 °C (non-condensing)
  • Operating Altitude: Up to 2,000 m
  • Non-operating Altitude: Up to 4,600 m

Note: The G1310B is an isocratic pump with a single solvent channel and does not include an integrated degasser. An external vacuum degasser is required for operation at flow rates below 0.5 mL/min.