The Agilent Technologies 1260 Infinity Isocratic Pump (G1310B) is an isocratic HPLC pump designed for precise and stable mobile phase delivery. It features a dual-piston, in-series pumping system with a servo-controlled variable stroke drive. This design provides consistent solvent flow, high precision, and low pressure pulsation.
The pump offers a flow range of 0.2–10.0 mL/min and supports operating pressures up to 600 bar. Its features include adjustable mobile phase compressibility compensation, integrated diagnostics, and leak detection. The Early Maintenance Feedback (EMF) function helps monitor instrument usage and plan maintenance activities. The G1310B is suitable for isocratic HPLC separations requiring a constant mobile phase composition.
Technical Specifications
General Specifications
- Model: Agilent 1260 Infinity Isocratic Pump (G1310B)
- Technology: Dual-piston, in-series pump with servo-controlled variable stroke drive
- Flow Range: 0.2–10.0 mL/min
- Flow Precision: ≤0.07% RSD or ≤0.02 min SD (whichever is greater)
- Flow Accuracy: ±1% or 10 µL/min (whichever is greater)
- Operating Pressure:
- Up to 60 MPa (600 bar) at flow rates up to 5 mL/min
- Up to 20 MPa (200 bar) at flow rates up to 10 mL/min
- Pressure Pulsation: < 2% amplitude (typically < 1.3%) or < 0.3 MPa (3 bar), whichever is greater, measured at 1 mL/min with isopropanol at pressures above 10 bar
- Compressibility Compensation: User-selectable, based on mobile phase compressibility
Physical and Electrical Characteristics
- Weight: 11 kg (25 lbs)
- Dimensions (H × W × D): 180 × 345 × 435 mm
- Voltage: 100–240 VAC, ±10%
- Frequency: 50/60 Hz, ±5%
- Maximum Power Consumption: 180 VA, 55 W / 188 BTU
Connectivity & Control
- Communication Interfaces:
- CAN
- RS-232C
- APG Remote (ready, start, stop, shutdown signals)
- optional LAN
- Software Control:
- Agilent ChemStation
- EZChrom
- MassHunter
- OpenLab
- Local Control: Agilent Instant Pilot
- Analog Output: 1.33 mV/bar for pressure monitoring
Maintenance & Safety Features
- Diagnostics: Error detection and monitoring through Agilent LabAdvisor
- Leak Detection: Leak detection, safe leak handling, and shutdown signaling
- Maintenance Monitoring: Early Maintenance Feedback (EMF) for tracking instrument usage, seal wear, and pumped mobile phase volume
- Safety Standards: IEC, CSA, UL; Installation Category II, Pollution Degree 2 (indoor use only)
Environmental Conditions
- Operating Temperature: 4–55 °C
- Non-operating Temperature: -40 to 70 °C
- Humidity: < 95% RH at 40 °C (non-condensing)
- Operating Altitude: Up to 2,000 m
- Non-operating Altitude: Up to 4,600 m
Note: The G1310B is an isocratic pump with a single solvent channel and does not include an integrated degasser. An external vacuum degasser is required for operation at flow rates below 0.5 mL/min.