Waters Arc Premier UPLC System (QP, Sampler, TCC, UV-Vis, PC, SW) (Refurbished)

Waters Arc Premier with QSM-R, SM FTN-R, Column Heater, and 2489 UV/Vis provides 655 bar operation, quaternary gradients, and MaxPeak HPS fluidics. The offered system also includes a PC, software, and licenses.
Rental per month
1,500
€
without VAT
Purchase price
42,500
€
without VAT
In stock
+ 500 € delivery

Features

Condition:
Refurbished
Warranty:
3 months
Manufacture year:
2021
Countries:
Czech Republic
Additional services
Installation
Advanced training
Extended warranty
Basic training
Application implementation

Features

The Waters Arc™ Premier System is a high-performance liquid chromatography platform designed for robust HPLC/UHPLC separations, method development, method transfer, and routine QA/QC workflows. The offered configuration includes a Quaternary Solvent Manager, Sample Manager FTN, column heater, Arc Premier 2489 UV/Vis Detector, PC workstation, software, and licenses.

A key feature is MaxPeak™ High Performance Surfaces (HPS) Technology, which minimizes nonspecific interactions between metal-sensitive analytes and metallic surfaces in the sample flow path. This can improve recovery, peak shape, reproducibility, and overall data quality for compounds prone to metal interactions.

Offered System Configuration

  • Waters Arc Premier System
  • Quaternary Solvent Manager-R (QSM-R)
  • Sample Manager FTN-R
  • Column Heater
  • Arc Premier 2489 UV/Visible Detector
  • PC / workstation
  • Software and licenses

Quaternary Solvent Manager-R

The QSM-R is a low-pressure-mixing, high-pressure quaternary solvent delivery module capable of dynamically blending four degassed solvents.

QSM-R Specifications
  • Gradient: Quaternary low-pressure mixing
  • Solvent channels: 4
  • Flow range: 0.001–5.000 mL/min
  • Maximum pressure: 655 bar / 9500 psi
  • Standard mixer: 340 µL titanium
  • Integrated degassing: four channels
  • Flow accuracy: ±1.0% under specified conditions
  • Flow precision: 0.05% RSD or 0.01 min SD
  • Composition accuracy: ±0.5% absolute
  • Composition precision: 0.15% RSD or 0.02 min SD
  • Leak sensing
  • Programmable seal wash
  • Automatic Prime
  • Flow Ramping
  • Auto•Blend Plus.

Sample Manager FTN-R

The SM FTN-R uses a flow-through-needle mechanism in which the sample is aspirated into the needle and the needle becomes part of the high-pressure flow path during injection.

SM FTN-R Specifications
  • Standard injection range: 0.1–50 µL
  • Optional range: up to 0.1–1000 µL
  • Standard extension loop: 50 µL
  • Maximum operating pressure: 655 bar / 9500 psi
  • Flow range: 0–5 mL/min
  • Sample temperature range: 4–40 °C
  • Temperature increment: 0.1 °C
  • Heating: up to approx. 25 °C above ambient
  • Cooling: up to approx. 19 °C below ambient
  • Temperature sensor accuracy: ±0.5 °C
  • Temperature stability: ±1.0 °C
  • Residual sample volume: 3 µL with Total Recovery vial
  • UV carryover: <0.002% caffeine under specified test conditions.
Injection Reproducibility
  • ≤1% RSD at 0.5–0.9 µL
  • ≤0.5% RSD at 1.0–4.9 µL
  • ≤0.25% RSD at ≥5.0 µL.

Column Heater

For the standard Column Heater-Active (CH-A) configuration:

  • One column
  • Maximum column ID: 4.6 mm
  • Maximum length: 150 mm including filter or guard column
  • Temperature range: 20–90 °C
  • Temperature accuracy: ±0.5 °C
  • Temperature stability: ±0.3 °C
  • Active preheating standard
  • Optional passive preheating
  • eCord™ column tracking.

If the offered system instead includes the CH-30A, columns up to 300 mm can be accommodated.

Arc Premier 2489 UV/Visible Detector

The Arc Premier 2489 UV/Vis Detector is a dual-wavelength absorbance detector incorporating MaxPeak HPS technology in the inlet and outlet fluidics of the flow cell.

2489 UV/Vis Specifications
  • Detection: Dual-wavelength UV/Vis
  • Wavelength range: 190–700 nm
  • Bandwidth: ≤5 nm
  • Wavelength accuracy: ±1 nm
  • Wavelength repeatability: ±0.1 nm
  • Measurement range: 0.0001–4.0000 AU
  • Sampling rate: up to 80 points/s
  • Single-wavelength baseline noise: ≤5 × 10⁻⁶ AU
  • Drift: ≤1 × 10⁻⁴ AU/hour
  • Light source: Deuterium arc lamp
  • Analytical path length: 10 mm
  • Analytical cell volume: 16.3 µL
  • Flow-cell pressure limit: 1000 psi
  • Flow-cell design: TaperSlit™
  • MaxPeak HPS inlet/outlet fluidics.

Overall System Performance

For the QSM-R configuration:

  • Total system bandspread (4σ): ≤30 µL
  • Total dwell volume: ≤1150 µL
  • Gradient delay volume: ≤750 µL
  • Flow range: 0.001–5.000 mL/min
  • Maximum operating pressure: 655 bar / 9500 psi
  • Injection-to-injection cycle time: ≤30 s.

Typical Applications

  • Pharmaceutical development
  • QA/QC
  • HPLC method transfer
  • UHPLC analysis
  • Impurity profiling
  • Assay
  • Metal-sensitive analytes
  • Phosphorylated compounds
  • Nucleotides
  • Organic acids
  • Food analysis
  • Environmental analysis
  • Chemical materials
  • Routine UV/Vis quantitation