The Agilent 1260 Infinity GPC/SEC Multi-Detector Suite (G7800A) is an advanced multidetector platform for gel permeation chromatography and size exclusion chromatography (GPC/SEC), optimized primarily for polymer characterization in organic solvents. The offered configuration combines three complementary detection principles—differential refractive index (RI), dual-angle light scattering (LS), and viscometry—providing significantly more structural information than conventional column-calibrated GPC alone.
Combining RI, LS, and viscometry enables determination of polymer concentration, absolute molecular weight, and information about molecular size, conformation, and branching. Each detector is independently temperature controlled from 30 to 60 °C for stable baselines and reproducible measurements. The offered system also includes two Agilent PLgel 5 µm MIXED-C columns for broad-range polymer separations.
Offered Configuration
G7801A – Refractive Index Detector
The differential RI detector acts as the universal concentration detector in the GPC/SEC workflow.
- Detection principle: Differential refractive index
- Detection type: Deflection
- Wavelength: 658 nm
- Cell volume: 6 µL
- Short-term noise: <80 µV
- Pressure rating: 350 kPa / 50 psi
- Temperature range: 30–60 °C
- Temperature stability: ±0.2 °C
- pH range: 2–10
- Wetted materials: 316 SST, quartz
G7803A – Dual Angle Light Scattering Detector
The dual-angle LS detector measures polymer molecular weight and molecular dimensions in solution without requiring conventional column calibration to assign molecular weight.
- Detection: Dual Angle Static Light Scattering
- Scattering angles: 15° and 90°
- Laser wavelength: 658 nm
- Sample cell volume: 10 µL
- Scattering volume: 0.01 µL
- Temperature range: 30–60 °C
- Temperature stability: ±0.2 °C
G7802A – Viscometer
The integrated viscometer uses a balanced four-capillary bridge and measures pressure changes created by polymer solutions flowing through the detector.
- Principle: Four-capillary differential viscometer
- Typical shear rate: 3000 s⁻¹
- Sensitivity ηsp: 1 × 10⁻⁵ Pa·s
- Pressure rating: 50 kPa / 7 psi
- Recommended maximum flow: 2 mL/min
- Temperature range: 30–60 °C
- Temperature stability: ±0.2 °C
- pH range: 2–10
PLgel 5 µm MIXED-C Columns
The supplied system includes 2× Agilent PLgel 5 µm MIXED-C columns.
- Particle size: 5 µm
- Particle type: mixed-pore, fully porous
- Standard dimensions: 7.5 × 300 mm
- Molecular weight range: 200–2,000,000 Da (PS equivalent)
- Maximum pressure: 150 bar
- Maximum temperature: 150 °C
- Typical operating flow: approximately 1.0 mL/min
- Maximum flow: 1.5 mL/min
- Recommended configuration: 2 × 300 mm
- Typical applications include polystyrenes, polyurethanes, polycarbonates, and polysiloxanes.
G7800A Technical Specifications
- Type: Multidetector GPC/SEC platform
- Offered Detector Configuration: RI + Dual Angle LS + Viscometer
- Enclosure: Quad Compartment
- Quad Compartment Dimensions: 400 × 490 × 650 mm
- Quad Compartment Weight: approximately 20 kg
- Line Voltage: 100–240 VAC ±10%
- Frequency: 50/60 Hz
- Maximum Power Consumption: 150 W
- Ambient Operating Temperature: 15–30 °C
- Humidity: <95% at 30 °C, noncondensing
- Maximum Operating Altitude: 2000 m
Typical Applications
- Gel permeation chromatography
- Size exclusion chromatography
- Polymer molecular weight characterization
- Molecular weight distribution analysis
- Absolute molecular weight by light scattering
- Intrinsic viscosity measurement
- Polymer branching analysis
- Polymer conformation and structure studies
- Polymer and materials research
- Polymer quality control
- Resins, plastics, elastomers, and other macromolecules