Agilent 1200 High Performance Autosampler SL, G1367C (Refurbished)

The Agilent G1367C is particularly useful for laboratories that routinely switch between vials, tubes, and well plates while requiring long unattended sequences. Capacity for up to two 384-well plates combines high sample throughput with 600 bar LC operation.
Purchase price
4,500
without VAT
In stock
+ 150  delivery

Features

Condition:
Refurbished
Warranty:
3 months
Manufacture year:
0
Countries:
Czech Republic
Additional services
Installation
Extended warranty
Basic training

Features

The Agilent 1200 G1367C High Performance Autosampler SL is a flexible automated injector for laboratories working with multiple sample-container formats. It supports standard vials, Eppendorf tubes, and microplates, and two 384-well plates provide capacity for up to 768 unattended samples. Combined with a 600 bar pressure capability, this flexibility makes the sampler useful for both conventional and high-speed LC workflows.

Technical Specifications:

  • Injection Volume Range: 0.1 – 100 µL in 0.1 µL increments; up to 1500 µL with multiple draw (hardware modification required).
  • Injection Precision:
    • < 0.25% RSD for 5 – 100 µL.
    • < 1% RSD for 1 – 5 µL.
  • Pressure Range: 0 – 600 bar (0 – 8700 psi).
  • Sample Viscosity Range: 0.2 – 5 cP.
  • Sample Capacity:
    • 2 microplates + 10 × 2 mL vials
    • up to 108 × 2 mL vials in two 54-position trays plus 10 additional vials
    • 30 × 6 mL vials in two 15-position trays
    • Eppendorf tubes with appropriate trays
    • 96- and 384-well microplates​
  • Injection Cycle Time: Typically < 17 seconds using standard conditions (default draw speed: 100 µL/min; default eject speed: 100 µL/min; injection volume: 1 µL).
  • Carryover: Typically < 0.01% under specified conditions.
  • Communications
    • CAN
    • RS-232C
    • APG Remote​
  • Safety Features: Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display.
  • GLP Features: Early maintenance feedback (EMF), electronic records of maintenance and errors.
Physical Specifications
  • Dimensions (H × W × D): approximately 200 × 345 × 440 mm
  • Weight: approximately 15.5 kg
  • Power Supply: 100–240 VAC
  • Frequency: 50/60 Hz
  • Maximum Power Consumption: 200 W
  • Ambient Operating Temperature: 4–55 °C